Automated Electrophysiology Assays

, , , , , , and .

Author Information and Affiliations

Published .

Introduction

Technological breakthroughs in electrophysiological techniques, such as development of voltage-clamp and patch clamp, have enabled direct measurements of membrane currents in single cells. The low throughput of traditional electrophysiology, however has limited its use in the drug discovery process. High-throughput, non-electrophysiology based primary assays were used successfully but lack key information content, such as voltage dependence and kinetics, and may not be suitable for all ion channels. Most importantly, non-electrophysiological assays lack the ability to accurately control membrane potential, preventing the selective targeting of disease relevant conformational states of voltage-gated channels. For these reasons, substantial efforts were devoted to the development of instruments capable of automated electrophysiological recording at moderate- to high-throughput.

Early on, a number of different approaches were investigated (1). The field quickly settled on planar substrates, and in 2002 two automated electrophysiology instruments came on the market: PatchXpress® and IonWorks™ HT developed at Axon Instruments and Essen BioSciences, respectively, and later sold to Molecular Devices. Several other manufacturers launched their automated instruments in subsequent years; most notably Sophion (QPatch), Nanion (Patchliner®) and Fluxion (IonFlux). Based on learnings obtained with first generation devices, a new set of second generation instruments was developed in more recent years.

Instruments differ substantially in the composition of the patch plate, the configuration of the electrodes, liquid handling methods, and the software used to control the experiment and analyze data, and these features are discussed in greater detail below. Each platform has its strengths and weaknesses, and consideration should be given to the intended purpose before a device is purchased.

Flowchart

The flowchart for automated electrophysiology assays comprises three sections: requirements for assay protocol, choice of instrumentation, and choice of cell line. The first section, requirements for assay protocol, includes a number of important considerations beginning with how the target channel is activated (constitutively active, ligand-gated or voltage-gated). The second consideration is whether the indication requires an agonist, blocker or modulator. The third consideration is if multiple additions and/or wash-outs are needed for the protocol. The fourth consideration is whether data will need to be acquired during compound addition. The fifth consideration is the requirements for sampling frequency, and lengths and complexity of the voltage protocol. Finally one needs to consider the throughput that will be required. Within the second section a number of important considerations for the choice of instrumentation are listed. One needs to consider whether low-throughput (single cell), medium-throughput (8-16 well) or high-throughput (96- or 384-well) formats are needed. Secondly, one needs to be aware that a discontinuous voltage clamp limits the suitability of some platforms for ligand-gated and other channels. It is also important to consider needs for a current clamp and temperature control, as well as the suitability of seal quality, fluid handling and amenable voltage protocols. The final section is choice of cell line, which can be influenced by multiple considerations. The typical cell requirements are about 10 million cells in 96- or 384-well formats. Recombinant cell lines are often utilized for electrophysiology assays and CHO, CHL and HEK lines have all been used successfully. Some native cell preparations and primary lines have also been used, such as T-lymphocytes and RBL-1 cells. The ability to grow cells in suspension is advantageous for high-throughput applications.

Choice of Instrumentation

Several excellent reviews of automated electrophysiology have been published (2-5) and may be consulted for a more detailed description of the available technologies. A general overview follows below.

Manual electrophysiology is enabled by the formation of a high resistance electrical seal between cell membranes and the glass used to manufacture patch pipettes. Most of the early attempts at automating electrophysiological recordings tried to replace the glass pipette with small holes in planar glass surfaces. Some medium-throughput instruments (e.g. PatchXpress) still use this approach. Recordings on these instruments typically achieve high resistance gigaohm (GΩ) seals, similar to manual electrophysiology. However, from a manufacturing standpoint, producing uniform holes of 1-2 μm diameter in planar glass represents a significant challenge. Attempts at increasing the throughput by increasing the number of parallel recordings or decreasing variability through ensemble recordings therefore employ alternative substrates. The consequence of using plastic substrates, however, is a lower seal resistance. For many applications, seal resistances in the 100-200 megaohm (MΩ) range are sufficient to produce reliable data. Limitations may still exist for currents with low amplitude, rapid activation kinetics or very steep voltage dependence, or for currents lacking voltage-dependent gating where the channel current may be difficult to distinguish from background leak currents. Small currents may result in a poor signal-to-noise (S/N) ratio; and insufficient response speed and accuracy of voltage control may compromise accurate measurement of rapidly activating, steeply voltage-dependent currents.

Most platforms now offer the option of recording an “ensemble” current, referred to as population patch clamp (PPC) (6). In this configuration, each well in a seal chip/patch plate contains multiple (up to 64) holes for recording of an integrated total ensemble of currents across a number of individual cells. The advantage is greater uniformity, which typically equates to higher success rates (often as high as 100%). The uniformity afforded by PPC enables endpoint assays in which currents from wells receiving the test article are compared directly to control wells, without the need for normalization to pre-compound currents.

A consequence of ensemble recordings is the requirement for much larger currents to be passed across the electrodes, particularly if seal resistances are relatively low. Large current amplitudes can contribute to electrode instability, especially on high-throughput platforms with shared ground electrodes such as IonWorks Barracuda®.

Problems with electrode instability are minimized by limiting the length of the voltage protocol, avoiding extreme voltage steps, especially in the hyperpolarizing direction, and reconditioning ground electrodes daily. For these reasons, early presentations on IonWorks Barracuda were often limited to recordings of ligand-gated channels, such as GABA-A channels. Including a silver chelator in the recording solutions may also help to avoid issues related to block of currents by silver ions emanating from the electrodes (7).

Most high-throughput electrophysiology assays for testing pharmacological agents are based on assessing whole cell currents, i.e. currents that flow across the entire cell membrane. The whole cell recording configuration can be established by two methods. Although both are technically whole cell recordings, the two methods are conventionally referred to as whole cell and perforated patch recordings, respectively. During whole cell recordings, the patch of membrane exposed to the intracellular solution, either in the patch pipette or across the hole in the planar seal chip/patch plate, is ruptured by applying negative pressure or a voltage transient. Typically the strength, duration and frequency of the negative pressure pulses can be varied and has to be optimized for each cell line such that good electrical access is achieved without damage to the cell or compromising the membrane-substrate seal. Good starting points for typically used cell lines (see below) can be obtained from the instrument manufacturers. Alternatively, perforated patch recordings establish electrical access by exposing the same membrane patch to agents that form non-selective cation pores within the lipid bilayer. More details on the choice of perforating agents follow below.

Choice of Cell Line

All automated electrophysiology instruments use a cell suspension that is pipetted into the recording chamber, where negative pressure facilitates the placement of single cells over the holes in the planar recording substrate. Requirements for cell numbers are typically on the order of 107 cells per 96- or 384-well recording. Given the requirement for preparing a cell suspension and for large numbers of cells per experiment, non-adherent cells offer advantages. However, historically, most ion channels have been stably expressed in either adherent Chinese Hamster Ovary (CHO) cells or in adherent Human Embryonic Kidney (HEK) cells, because of the limited expression of endogenous channels, the propensity to form high resistance electrical seals on a glass surface, and the widespread use of these host cells for optical assays. Therefore, automated electrophysiology instrumentation has been optimized to work with these parental cell types. Good success rates have also been reported with Chinese Hamster Lung (CHL) and RBL-1 cells. If expression in another cell background is desired, it would be advantageous to first test the sealing properties of the parental cells on the automated electrophysiology instrument that will be used with the cell line. For other considerations regarding cell preparation and for a detailed protocol for preparing CHO and HEK cell suspensions for use on automated electrophysiology instruments, see the chapter on Ion Channel Screening.

It is possible to use acutely dissociated native cells on automated electrophysiology devices; however, there are a number of important considerations and limitations (8). Cells have to be available in large numbers and the dissociation procedure has to produce a fairly homogeneous preparation of cells without a lot of cell debris. Because of these limitations, the native cell type used most successfully has been T-lymphocytes. Patchliner may offer an advantage for recording from native cells, because Nanion offers patch plates with different hole sizes and a low cell usage protocol, requiring as few as 300 cells per 16-well recording (9). Patchliner recordings from several native cell preparations have been reported. For some applications, induced pluripotent stem cells may offer a viable alternative to native cell preparations (10).

Choice of Solutions

The desire for higher resistance seals has led many researchers to employ intracellular solutions with high fluoride content. Intracellular solutions with F- as the major anion lead to dramatically increased measured seal resistances; however, this may not be associated with an increase in the quality of the recording and may have unintended consequences such as changes in intracellular signaling and phosphorylation. For example, time dependent changes in sodium channel biophysics in the presence of high intracellular F- were described in detail by Wang et al (11). These experiments were conducted using manual electrophysiology, but a similar lack of stability has been seen on the automated platforms. The reader is cautioned to explore alternatives and not to be persuaded by high resistance numbers alone.

High extracellular concentrations of divalent ion can also increase seal resistances, especially in the presence of high intracellular F-. It may be helpful to temporarily expose cells to high extracellular Ca2+ concentrations during the sealing process, and change to a more physiological solution once seals are established. This is an approach recommended for Patchliner, which was designed with a space and protocol for use of a “seal enhancer” solution. Much of the increased seal resistance achieved by temporary exposure to a high divalent extracellular solution is maintained once the divalent concentration is lowered. While it is possible on Patchliner to exchange the intracellular solution, lowering the intracellular F- concentration after seal formation leads to a dramatic loss of seal resistance (BTP, personal experience).

High-Throughput Automated Electrophysiology Assays

For the purpose of this discussion, high-throughput will be defined as the parallel recording from either 96 or 384 wells.

First Generation Instruments: IonWorks Quattro and IonWorks HT

IonWorks Quattro and IonWorks HT are distinguished by the use of single hole and PPC recording mode, respectively. Both instruments feature 48 movable electrodes and discontinuous voltage clamp and use the perforated patch configuration. The same voltage protocol is performed prior to and after a fixed incubation time with the test article.

Single Hole versus Population Patch Clamp (PPC)

On IonWorks instruments, cells can be supplied in a trough or in individual wells; this allows the user to test multiple different cell lines in the same experiment and makes the instruments well suited to selecting stable cell lines out of multiple clones. For this application, the Single Hole mode is ideal because it provides information on the distribution of current amplitudes across cells.

For other applications, the PPC mode is typically preferred (12). Because currents are effectively averaged across 64 cells, the PPC mode mitigates against cell-to-cell variability in parameters such as current amplitudes or fractional inactivation, and recordings are much more uniform. Often it is possible to achieve 100% success rate. However, if seal rates are poor, it may be possible to acquire data from a number of higher resistance wells in Single Hole mode, whereas recordings in PPC mode will fail altogether. For this reason, the Single Hole mode should be used to troubleshoot experiments when PPC recordings report “no seal” across the entire plate. If the reason for the failure is poor cell health, the Single Hole experiment should produce some wells with seal resistances that pass the intrinsic acceptance criteria (resistance >20 MΩ during seal test). If the Single Hole experiment shows no wells with acceptable seal resistances, the reason for the failure is most likely due to instrument failure or an error in setting up the experimental protocol.

Perforation Agent

The most commonly used perforation agent for IonWorks experiments is amphotericin B (from Streptomyces), although nystatin should be a viable alternative and is used more often for manual electrophysiology. There are important considerations when using amphotericin B. It is light sensitive and should be stored in the dark. Amphotericin B has limited solubility. It should always be prepared fresh on the day of the experiment, dissolved in 100% dry DMSO at approximately 30 mg/mL, sonicated for several minutes, and then diluted into the aqueous solution to a final concentration of 0.1 mg/mL. Amphotericin B of comparable quality is available from multiple vendors; however, lot differences exist and can affect various cell lines differently. Therefore it may be worthwhile to compare different lots of amphotericin B during assay development.

Although amphotericin B and nystatin form cation-selective pores, their permeability for chloride anions is not insignificant (13). This deserves some consideration when developing assays on the IonWorks platforms that employ the perforated patch method. In cases where the chloride concentration in the intracellular solution is high compared to typical cytoplasmic free chloride concentration, the cytoplasmic chloride concentration will most likely have reached steady-state at the time the actual data recording starts. However, this may not be the case; if the chloride concentration in the intracellular solution is low compared to the cytoplasmic free chloride concentration, this may lead to instability over time. Gramicidin may represent a viable alternative as a perforating agent and is much more selective for cationic currents (13).

Voltage Protocols

The complexity of voltage protocols that can be used on IonWorks HT or IonWorks Quattro is limited. The software is best suited for single voltage steps or ramps or trains of voltage pulses at a constant frequency. Generally, the length of the voltage protocol is limited to 5 s. However, another limitation exists in the number of data points that can be acquired, and therefore the maximal length of recordings depends on the sampling frequency employed. Recordings can be performed at sample intervals ranging from 0.1 ms to 6.4 ms. A sampling interval of 0.1 ms allows for voltage protocols lasting up to 2.25 s. It is possible to include conditioning intervals or trains, during which cells are voltage clamped but no data are acquired. Adding conditioning segments can quickly extend the total time of the experiment, because recordings are performed in 8 banks of 48 recordings, and it will be important to determine the stability of currents over the time course of the experiment.

Data Analysis

Leak subtraction is performed based on a small voltage step at the beginning of the voltage protocol. Because of the error in extrapolating from a small voltage step to the entire voltage range of the experiment and the possibility of offset potentials, the use of leak subtraction should be combined whenever possible with an offset correction based on leak currents recorded during a voltage step to 0 mV. However, this may not be possible if a membrane potential of 0 mV produces active currents. A good alternative for fast inactivating channels is the measurement of currents based on subtracting current at the end of a voltage step from the peak current.

For non-inactivating currents that operate across the voltage range, leak subtraction is not an option, and a combination of a voltage step to 0 mV to minimize the contribution of leak currents combined with a voltage ramp has been used successfully (14).

After data has been acquired, it can be filtered based on acceptance criteria such as seal resistance or current amplitude. The user then chooses one or several metrics that will be used to analyze the current traces such as the peak or average current during a particular interval. These data are stored in the form of ASCII files that can be imported into a variety of software programs for further data analysis. Overall, the analysis of IonWorks recordings is rapid and user friendly.

Example Protocol

See Appendix 1 for a detailed protocol that can be used to evaluate block of voltage-gated sodium currents on the IonWorks platform. Other voltage-gated currents can be recorded using the same workflow with appropriate adjustments in the voltage protocols and potentially in the composition of recording solutions.

Second Generation Instruments: IonWorks Barracuda, Qube and SyncroPatch

A key limitation of the first generation instruments is the inability to voltage clamp during test article addition. In this regard, second generation high-throughput instruments represent a significant advantage by enabling continuous voltage clamp. Another advantage of the 384 amplifiers is the shorter time required to conduct an experiment, i.e. the run time of a typical experiment is reduced from approximately one hour on IonWorks Quattro to 15-20 minutes on the second generation instruments. All platforms offer a choice between Single Hole and PPC modes, and the same considerations apply that are discussed above for IonWorks Quattro and IonWorks HT. IonWorks Barracuda and Qube (Sophion) are free-standing instruments, whereas SyncroPatch 384PE (Nanion Technologies) is designed to be integrated into a liquid handling platform, such as Biomek® FX (Beckman Coulter).

Whole Cell Access

Like the original IonWorks instruments, IonWorks Barracuda uses the perforated patch configuration, typically resulting in low access resistance that is very stable over time. In contrast, Qube and SyncroPatch obtain whole cell access by applying negative pressure to rupture the cell membrane. Different cell lines can differ substantially with regard to the negative pressure required to rupture the membrane, and the associated parameters need to be optimized carefully during assay development. In general, the user can vary the number, length and amplitude of the pressure pulses. Options may also include pressure ramps. If no or very little negative pressure is required for whole cell access, the cell membrane may have been damaged by excessive enzyme treatment and a gentler dissociation procedure should be tried.

Voltage Protocols

Second generation instruments can handle fairly complex voltage protocols, including protocols that combine different holding potentials and sampling rates. All instruments feature graphic user interfaces for setting up voltage protocols and experimental parameters. It is reasonable to assume that protocols commonly used to characterize voltage-gated channels can be implemented on these second generation instruments, and the reader is referred to the instrument manufacturer for questions concerning specific protocols.

Data Analysis

All second generation platforms include customized data analysis software that allow the user to subtract leak currents, set well-level criteria for data acceptance and define metrics that specify the data (such as peak current, average current during a specified time interval, etc.) that will be used for further analysis. Some software programs support further analysis of data across wells, such as IC50 or EC50 determination, whereas other platforms require well-level data to be exported and analyzed separately. Generally, the latter is not an impediment, because most users will already have access to analysis software that is integrated with compound management information and database publication tools. Nonetheless, real-time analysis can be a nice feature when deciding whether to repeat experiments.

Example Protocols

See Appendix 2 for a protocol that can be used to evaluate block of KCNQ1/KCNE1 voltage-gated potassium currents on the IonWorks Barracuda platform.

Appendix 3 contains an example of a protocol that can be used to record currents from ligand-gated ion channels on IonWorks Barracuda. The example given is for GABAA currents; similar protocols can be used with other ligand-gated chloride and cationic currents, although fluidic handling on the instrument precludes accurate recording of the most rapidly activating and inactivating channels (NMDA and nAChR α7).

Medium-Throughput Automated Electrophysiology Assays

Medium-throughput instruments use 8- or 16-well chambers. Currently available commercial instruments include PatchXpress, Patchliner and Qpatch. PatchXpress and Qpatch are free-standing instruments, whereas Patchliner is a bench-top instrument. A difference between these platforms is the configuration of the electrodes that can be either movable (PatchXpress and Patchliner) or embedded into the patch plate (Qpatch). Movable electrodes involve the risk of carryover of test articles between experiments, but are less prone to electrical cross-talk between electrodes, potentially leading to recording artifacts.

The PPC mode is available for Patchliner and Qube instruments.

Scripting

The PatchXpress and Patchliner instruments offer the option of scripting, referring to protocols that are adjusted ad hoc based on criteria applied to the data. A commonly used example is the addition of test article once currents evoked by baseline pulses have stabilized and vary by less than 5%. Another commonly used application of scripting is to set the test pulse voltage based on the maximum current determined during a family of voltage pulses. Especially for cell lines with low expression, this type of script can help maximize the number of successful recordings. Most modulators of voltage-gated sodium and calcium channels are sensitive to the conformational state of the channel and typically interact preferentially with channels in open and inactivated states. To avoid potency differences based on cell-to-cell variability in the voltage-dependence of inactivation, scripting can be used to set the membrane potential on the fly to a voltage corresponding to a set fractional inactivation based on steady state inactivation data.

Data Analysis

PatchXpress and Qpatch come with analysis software that is flexible but time consuming. Analysis performed by Patchliner software is limited to all-points IC50 values and more detailed analysis requires importing the data into another program.

Plate-based analysis software used by most pharmaceutical companies and contract research operations is incompatible with the data formats used by medium-throughput automated electrophysiology devices.

Example Protocol with Scripting

See Appendix 4 for a protocol designed to evaluate block of voltage-gated sodium currents on the PatchXpress.

Low-Throughput Automated Electrophysiology Assays

The Port-a-Patch offered by Nanion is comparable to manual electrophysiology in terms of throughput. Its advantages are the small footprint and ease of operation relative to manual electrophysiology. It may be a viable option for laboratories new to electrophysiology. The main disadvantage is that the experimenter does not choose the individual cells that will be voltage clamped. This limits the use with transiently transfected or acutely dissociated cells.

Cost of Operation

The cost of consumables for any automated electrophysiology platform can be substantial and should be considered prior to purchasing an instrument. Outside of the price per patch plate/seal chip, factors that affect the cost per data point relate to the options of multiple test article additions, wash-out of test article, cell reuse, and scripting.

References

1.
Mathes C. Ion channels in drug discovery and development. Drug Discov Today. 2003;8:1022–1024. [PubMed: 14690632]
2.
Wang X, Li M. Automated electrophysiology: high throughput of art. Assay Drug Dev Technol. 2003;1:695–708. [PubMed: 15090242]
3.
Dunlop J, Bowlby M, Peri R, et al. Ion channel screening. Comb Chem High Throughput Screen. 2008;11:514–522. [PubMed: 18694388]
4.
Dunlop J, Bowlby M, Peri R, Vasilyev D, Arias R. High-throughput electrophysiology: an emerging paradigm for ion-channel screening and physiology. Nat Rev Drug Discov. 2008;7:358–368. [PubMed: 18356919]
5.
Priest BT, Swensen AM, McManus OB. Automated electrophysiology in drug discovery. Curr Pharm Des. 2007;13:2325–2337. [PubMed: 17692004]
6.
Finkel A, Wittel A, Yang N, Handran S, Hughes J, Costantin J. Population patch clamp improves data consistency and success rates in the measurement of ionic currents. J Biomol Screen. 2006;11:488–496. [PubMed: 16760372]
7.
Cerne R, Wakulchik M, Li B, Burris KD, Priest BT. Optimization of a High-Throughput Assay for Calcium Channel Modulators on IonWorks Barracuda. Assay Drug Dev Technol. 2016;14:75–83. [PubMed: 26716356]
8.
Milligan CJ, Li J, Sukumar P, et al. Robotic multiwell planar patch-clamp for native and primary mammalian cells. Nat Protoc. 2009;4:244–255. [PMC free article: PMC2645065] [PubMed: 19197268]
9.
Becker N, Stoelzle S, Gopel S, et al. Minimized cell usage for stem cell-derived and primary cells on an automated patch clamp system. J Pharmacol Toxicol Methods. 2013;68:82–87. [PubMed: 23567076]
10.
Haythornthwaite A, Stoelzle S, Hasler A, et al. Characterizing human ion channels in induced pluripotent stem cell-derived neurons. J Biomol Screen. 2012;17:1264–1272. [PubMed: 22923790]
11.
Wang DW, George AL Jr, Bennett PB. Comparison of heterologously expressed human cardiac and skeletal muscle sodium channels. Biophys J. 1996;70:238–245. [PMC free article: PMC1224923] [PubMed: 8770201]
12.
Dale TJ, Townsend C, Hollands EC, Trezise DJ. Population patch clamp electrophysiology: a breakthrough technology for ion channel screening. Mol Biosyst. 2007;3:714–722. [PubMed: 17882333]
13.
Kyrozis A, Reichling DB. Perforated-patch recording with gramicidin avoids artifactual changes in intracellular chloride concentration. J Neurosci Methods. 1995;57:27–35. [PubMed: 7540702]
14.
Felix JP, Priest BT, Solly K, et al. The inwardly rectifying potassium channel Kir1.1: development of functional assays to identify and characterize channel inhibitors. Assay Drug Dev Technol. 2012;10:417–431. [PubMed: 22881347]

Appendix 1: Protocol for Assessment of Use-Dependent Sodium Channel Inhibitors on IonWorks Quattro

Overview

This protocol is designed to evaluate the type of use-dependent block of voltage-gated sodium channels that is seen with classic local anesthetics. After a brief period at a hyperpolarized voltage to allow recovery from inactivation, channels are subjected to a set of 26 depolarizing pulses. More pronounced inhibition of current during the later depolarizing pulses indicates use-dependent block, which typically arises from an increased affinity of test compounds for channels in open and inactivated states.

Start of Day

1.

Turn on IonWorks Quattro using the green button located on the front of the instrument on the lower right.

2.

Make sure waste bottles are empty.

3.

Make sure main vacuum is on and apply vacuum to patch plate in Patchplate holder slot (use an old patch plate for these steps).

4.

Place D-PBS without Ca2+/Mg2+ in all buffer solution bottles on the left side of the instrument (leave waste bottles empty).

Image aea-Image002.jpg

5.

Place a buffer trough in the Buffer position and fill with 5-7 mL of D-PBS without Ca2+/Mg2+.

Image aea-Image003.jpg

6.

After instrument initialization, start IonWorks software.

7.

From the Utilities menu, run the Flush & Rinse protocol twice.

Image aea-Image004.jpg

8.

Replace D-PBS with internal buffer solution (see below) in the bottles designated for the internal buffer solution and the antibiotic solutions. Make sure the correct lines are connected to each bottle.

9.

Replace the buffer in the buffer trough with 5-7 mL external buffer solution (see below).

10.

Run the Flush & Rinse protocol one time.

Buffer Solutions

External Buffer Solution: 140 mM NaCl, 4 mM KCl, 2 mM CaCl2, 1 mM MgCl2, 10 mM HEPES, 10 mM Glucose, pH 7.4 with NaOH

Internal Buffer Solution: 70 mM KF, 70 mM KCl, 0.5 mM MgCl2, 5 mM HEDTA, 10 mM HEPES, pH 7.25 with KOH

  • Amphotericin B is added to internal buffer solution to a final concentration of 0.1 mg/mL to make the antibiotic solution on the day of an experiment.
    i.

    Amphotericin B is weighed out and dissolved to a concentration of ~30 mg/mL in 100% anhydrous DMSO and then sonicated for 10 min.

    ii.

    Amphotericin B is then added to the internal buffer solution with vigorous mixing.

Protocol Setup

To load an existing protocol, select Open Protocol on the File menu.

Image aea-Image005

Once the protocol has loaded, it may be edited from the Edit menu by selecting Active Protocol.

Image aea-Image006

To set the location to save the data file, go to the Protocol Editor screen, select the File Saving tab and click the Browse button.

Image aea-Image007

The user can also select a file format. Options are a standard Date Prefix, a Custom Prefix, or customizing the file name after the protocol has started.

Image aea-Image008

On the Start >> Seal Test tab of the Protocol Editor, the user has to enter the type of patch plate (Standard or PPC) that will be used, as well as the time used for the seal test.

Image aea-Image009

To set up/edit the voltage protocol, select the Access >> End tab of the Protocol Editor and click the Voltage Protocol Editor.

Image aea-Image010

In the Voltage Protocol Editor, voltage protocols are generated by clicking Command Voltage and adding Interval Sequences. Each test pulse must be defined by an interval sequence. In the below example of a 200 ms step to -10 mV from -110 mV, the interval sequences consist of:

1.

Holding at -110 mV for 20 ms

2.

Holding at -10 mV for 200 ms

3.

Holding at -110 mV for 50 ms

By selecting the Apply repetition function, the user can repeat any of the interval sequences up to the memory capacity of the IonWorks program. In the below example, interval sequence 2 and 3 are repeated a total of 26 times.

Image aea-Image011

Also on the Access >> End tab, compound additions are controlled by using the Scan Parameters Editor.

Image aea-Image012

Start of Experiment

1.

Place compound plate in Plate 1 position (for guidelines on preparing compounds, see Compound Preparation below).

Image aea-Image013.jpg

2.

In Plate 2 position, place a polypropylene 384-well plate with 50 µl per well of a 50% DMSO/water mixture to wash F-head in between compound additions.

3.

In the Buffer position, place a buffer trough filled with fresh 5-7 mL of external buffer solution.

4.

Remove patch plate used during the Flush & Rinse protocols. With a piece of lint-free tissue, carefully dry out the Patchplate area, making sure not to touch the ground electrode.

5.

Place a new patch plate into the Patchplate area, replace the aluminum lid, and apply the vacuum. You should not be able to lift the lid after the vacuum is applied.

6.

Follow Cell Preparation directions below and place cells into the instrument.

7.

From main menu, click the green traffic light button to start program.

Image aea-Image014.jpg

Cell Preparation

1.

Obtain one 70-80% confluent T-75 flask of HEK cells stable expressing the Nav1 channel of interest.

2.

Aspirate media from flask.

3.

Add 10 mL D-PBS without Ca2+/Mg2+, tilt gently to wash the cell monolayer.

4.

Remove the D-PBS wash and add 2 mL trypsin (0.25%), tilt gently to wet the cell monolayer.

5.

Place flask in 37°C, 5% CO2 incubator undisturbed until cells start to round up and dislodge.

6.

Tilt flask to dislodge cells and/or tap sides of flask.

7.

Add 10 mL of cell media containing 10% FBS, washing the cells down the wall of the flask.

8.

Gently triturate the cells against the bottom of the flask four times.

9.

Place cell suspension into a 15 mL conical centrifuge tube and centrifuge @ 1300 rpm for 3 min.

10.

Aspirate supernatant and resuspend cells in 4 mL external buffer solution.

11.

Gently triturate the cells using a 200 µl pipette to break up the cell pellet. It is important that cell dispersion is carefully performed.

12.

Perform cell count.

13.

Cell concentration should be 2-2.5M cells/mL.

14.

Dilute with external buffer solution if needed.

15.

Carefully add entire contents of tube to trough located in Cells slot.

Compound Preparation

For concentration response curves (CRC), create a DMSO dilution plate in a 384-well plate following the plate map below to minimize effects from potential carryover of compounds on the E-head pins.

Image aea-Image015

Numbers listed in the plate map correspond to individual test compounds. Concentrations for each test compound should start with the lowest concentration in row A or C and go up in concentration as one moves down the plate (i.e. the highest concentration should be in row N or P).

For single point testing, compounds can be placed in any order.

Data Analysis

During an experiment, the IonWorks software generates an ASCII file based on one or more user defined analysis metrics. The metrics are defined under the Analyze/Export Data tab in the Protocol Editor. Examples of simple metrics, called Simple (Single Reduction) include peak current within a voltage interval and average current during a user defined time interval. Simple calculations, such as the difference or ratio between two currents, can also be applied to the data by selecting Compound (Two Reductions).

Image aea-Image016

Nav1 current traces recorded on IonWorks Quattro are shown below before (blue) and after (red) addition of 3 μM tetracaine. For the protocol described, a typical concentration-response curve (CRC) is shown for tetracaine. To generate this CRC, block during the last depolarization was determined at the well level by comparing Nav1 current amplitude before and after compound application, and data were exported out of the IonWorks software as % block, defined as (current in compound – current in control)/current in control x 100. Data were further analyzed by fitting a standard Hill equation to the data using Excel Fit.

Image aea-Image017

End of Day

1.

Replace external and internal buffer solutions with 70% ethanol.

2.

Place a buffer trough in the Buffer position and fill with 5-7 mL of 70% ethanol.

3.

On the Utilities menu, run Flush & Rinse protocol once.

4.

Close IonWorks Software and turn off IonWorks Quattro.

Appendix 2: Protocol for Assessment of KCNQ1/KCNE1 Currents on IonWorks Barracuda

Overview

This protocol is designed to evaluate inhibition of the voltage-gated potassium channel KCNQ1/KCNE1 during 3 s depolarizing test pulses repeated under control conditions and after addition of test compound.

Buffer Solutions (prepared in advance and stored at 4°C)

External Buffer Solution: 140 mM NaCl, 5 mM KCl, 2 mM CaCl2, 1 mM MgCl2, 10 mM HEPES, 10 mM Glucose, pH 7.4 with NaOH. Osmolarity 310-320 mOsm

Internal Buffer Solution: 90 mM KGluconate, 40 mM KCl, 3.2 mM MgCl2, 3.2 mM EDTA, 5 mM HEPES, pH 7.2 with KOH. Osmolarity 290-300 mOsm

  • Amphotericin B is added to internal buffer solution to a final concentration of 0.1 mg/mL to make the antibiotic solution on the day of an experiment.
    i.

    Amphotericin B is weighed out and dissolved to a concentration of ~30 mg/mL in 100% anhydrous DMSO and then sonicated for 10 min.

    ii.

    Amphotericin B is then added to the internal buffer solution with vigorous mixing.

Building Protocols

Setup Protocol

On the Edit Setup Protocol screen, input the following criteria into each of the different folder tabs:

Image aea-Image018
Image aea-Image019

Channel Protocol

On the Edit Channel Protocol screen, input the following criteria into each of the different folder tabs:

Step 1 specifies the scan and compound addition parameters. Please note the Advanced aspirate options as indicated.

Image aea-Image020

Step 2 specifies the pre-signal criteria.

Image aea-Image021

Step 3 specifies the conditioning train criteria. Please note that a conditioning train is not utilized for this particular protocol example.

Image aea-Image022

Step 4 specifies the different command voltage settings available.

Image aea-Image023

Step 5 specifies the sweep repetitions and timeline display of the command voltage protocol (by sweep or continuous display).

Image aea-Image024

Step 6 specifies the metric definitions. Please note that automatic export of metrics is turned off so as to allow manual inspection and rejection/acceptance of well traces before committing to metrics export.

Image aea-Image025

Cleanup Protocol

On the Edit Cleanup Protocol screen, input the following criteria into each of the different folder tabs:

Image aea-Image026

Running Experiments

Start of Day

1.

Turn on IonWorks Barracuda Instrument/computer using the switch located on right side of instrument.

2.

Open the IonWorks Barracuda Control Software.

3.

The instrument will go through an initialization.

4.

Make sure all waste containers are empty.

5.

Make sure there is 0.9% NaCl in both A & B wash containers.

6.

External solution bottle should be filled with DPBS (not external solution) that does not contain Mg and Ca.

7.

On the Utilities menu, select Load Cell Pipettor and follow the prompts.

8.

Fill both amphotericin B and Internal Solution bottles with internal solution containing 0.1 mg/mL amphotericin B.

9.

Place Model Cell on support pins above wash station.

10.

Insert ground electrodes into plenum (they must have soaked for at least 20 min in internal solution).

11.

Make sure a blank patch plate is inserted over the plenum and click Start of Day Flush from the Utilities menu.

Cell Preparation

1.

Start with >70% confluent T-150 flask of CHO cells stably expressing KCNQ1/KCNE1.

2.

Aspirate media, rinse with DPBS, and add 3 mL TrypLE™.

3.

Incubate 8 min at 37°C.

4.

Resuspend in 9 mL of growth media lacking selection factors.

5.

Spin cells at 1000 rpm for 3 min and resuspend in external solution at a density of 2-3 million cells/mL. A minimum of 5 mL of cell suspension are required.

6.

Place cells on IonWorks Barracuda.

Start of Experiment

1.

In Set Default Folders window, select the folders where the data and protocol files will reside.

Image aea-Image027.jpg

2.

Select the desired Setup, Channel, and Cleanup Protocols (as previously described).

3.

Replace patch plate used for Start of Day protocol with a fresh patch plate.

4.

Replace Model Cell with proper recording E-Head.

5.

Make sure tips, buffer boats, and compound plates are seated correctly on the instrument deck.

6.

Make sure cells and empty cell dispense plate are seated correctly.

7.

Set the experiment name.

8.

Click the green Play button to start the experiment.

Data Analysis

1.

From the Data Analysis button on the left, see below for an example of all the tabs that can be opened to configure data analysis and export.

Image aea-Image028.jpg

2.

On the Define Filters tab, set your desired hole/seal resistance, peak current amplitude, and/or other criteria for removing well data from the analysis.

3.

Open up the Metrics Editor and set the desired time and value measurements (an example is below):

Image aea-Image029.jpg

4.

Click the Metric Exports button on the left and define how and which metrics should be exported.

Image aea-Image030.jpg

KCNQ1/KCNE1 current traces recorded on IonWorks Barracuda are shown below before (black) and after (red) addition of XE-991. For the protocol described, a typical concentration-response curve (CRC) is shown for XE-991. Percent inhibition was calculated by averaging the last 10 ms of the current traces during the baseline depolarizations and during the last depolarizing step following the compound addition. Data were further analyzed by fitting a standard Hill equation to the data using Excel Fit.

Image aea-Image031

End of Day Flush

1.

Make certain 50% EtOH is in the alcohol wash bottle.

2.

Disconnect external solution bottle, amphotericin B bottle, and internal solution bottle and replace with empty 50 mL conical vials.

3.

Run End of Day protocol from the Utilities menu.

4.

Remove ground electrodes and place them back into internal solution soak.

5.

Shut everything off.

Appendix 3: Protocol for Assessment of GABAA Current Block on IonWorks Barracuda

Overview

This protocol is designed to evaluate inhibition of GABA-induced currents in a two-addition protocol. Small modifications allow this protocol to be used for agonists or allosteric modulators.

Buffer Solutions (prepared in advance and stored at 4°C)

External Buffer Solution: 140 mM NaCl, 5 mM KCl, 2 mM CaCl2, 1 mM MgCl2, 10 mM HEPES, 10 mM Glucose, pH 7.4 with NaOH. Osmolarity 310-320 mOsm

Internal Buffer Solution: 90 mM KGluconate, 40 mM KCl, 3.2 mM MgCl2, 3.2 mM EDTA, 5 mM HEPES, pH 7.2 with KOH. Osmolarity 265-275 mOsm

  • Amphotericin B is added to internal buffer solution to a final concentration of 0.1 mg/mL to make the antibiotic solution on the day of an experiment.
    i.

    Amphotericin B is weighed out and dissolved to a concentration of ~30 mg/mL in 100% anhydrous DMSO and then sonicated for 10 min.

    ii.

    Amphotericin B is then added to the internal buffer solution with vigorous mixing.

Cell Suspension (prepared immediately prior to experimental run)

1.

Start with 80-90% confluent T-150 flask of HEK cells stably expressing GABAA (same protocol is used for all GABAA isotypes).

2.

Aspirate media, rinse with DPBS, and add 3 mL TrypLE.

3.

Incubate 8 min at 37°C.

4.

Transfer to 15 mL conical tube, add 9 mL of growth media lacking selection factors and triturate (7-10 times).

5.

Spin cells at 1000 rpm for 3 min, aspirate supernatant and resuspend in external solution at a density of 3 million cells/mL.

6.

There should be at least 5 mL of cell suspension ready to be placed on IonWorks Barracuda.

Test Compounds

1.

Prepare a Mother Plate:

a.

Plate 50 μL of sample dissolved in 100% DMSO in columns 3 and 13 of 384-well plate.

b.

Serially dilute (10 point CRC with 1:3 steps) using liquid handler robot.

c.

Spin plate at 600 rpm for 1 min to collect all solution in the bottom of the well.

d.

Seal under argon and store at room temperature for up to one month (may not be suitable for all samples).

2.

Prepare Daughter Plates. On the day of experiment prepare compound and agonist plates:

a.

Dispense 100 μL of external solution into each well of a 384-well plate (Compound Plate).

b.

Dispense 100 μL of 1mM GABA containing external solution into each well of a 384-well plate (Agonist Plate).

c.

Add positive and negative controls to columns 1, 2, 23 and 24 of the compound plate.

d.

Immediately prior to the experiment, transfer 1 μL of DMSO stock solution from Mother Plate to Compound Plate and to Agonist Plate.

e.

Mix three times and spin plate at 600 rpm for 1 min.

Start IonWorks Barracuda (Start of Day)

1.

Turn on IonWorks Barracuda instrument/computer using the switch located on the right side of instrument.

2.

Open IonWorks Barracuda control software (the instrument will initialize automatically).

3.

Make sure all waste containers are empty.

4.

Make sure there is 0.9% NaCl in both A & B wash containers (approximately 4 L).

5.

External solution bottle should be filled with DPBS without Mg2+ and Ca2+ (prevents deposits).

6.

On the Utilities menu, select Load Cell Pipettor and follow the prompts:

Image aea-Image032.jpg

7.

Fill both amphotericin B and Internal Solution bottles with internal solution containing 0.1 mg/mL amphotericin B.

8.

Place Model Cell on support pins above wash station.

9.

Insert ground electrodes into plenum (they must have soaked for at least 20 min in internal solution).

10.

Make sure that a blank (uncontaminated with compounds) patch plate is correctly positioned in the Patch Plate position on the plenum.

11.

On the Utilities menu, select Start of Day Flush.

Select Data and Protocol Folders

On the File menu, select Set Data Folders:

Image aea-Image033

Select IW-Barracuda Protocols for GABAA

On the Protocol window select Setup, Command Voltage and Cleanup protocols:

Setup Protocol

From the Setup Protocol menu, input the following criteria into each of the different folder tabs:

Image aea-Image034
Image aea-Image019
1.

In the Add Cells panel select Wash before cell addition and specify the wash criteria

2.

In case that amphotericin B is not added to Internal Solution bottle, then Create Cell Access needs to be modified by the addition of Wait 240 s following the Add Perforation Agent

Channel Protocol

From the drop down “Channel Protocol” menu, select appropriate protocol.

1.

First compound addition:

a.

Set holding potential to -80 mV.

b.

Aspirate 10 μL of solution from Compound Plate (position Plate 2). Dispense at medium height at 5 μL/s.

c.

Record current 2 s prior and 30 s after compound addition with sampling frequency at 1 kHz.

d.

Follow with 3 min compound incubation.

e.

No scans are recorded prior or after compound addition.

Image aea-Image035.jpg

2.

Second compound addition:

a.

Maintain holding potential at -80 mV

b.

Aspirate 10 μL of solution from “Agonist Plate” (position Plate 1). Dispense at medium height at 5 μL/s

c.

Record current 2 s prior and 30 s after compound addition with sampling frequency at 1 kHz

d.

No scans are recorded prior or after compound addition

Image aea-Image036.jpg

For both, first and second dispense, adjust the Advanced Aspirate Option to the tip height of 20 μL, speed of 10 μL/s and extra volume to 2 μL. This setting minimizes the introduction of bubbles on dispense.

Image aea-Image037.jpg

3.

Set the resistance pulse (pre-signal):

Image aea-Image038.jpg

4.

Set the metric parameters:

Image aea-Image039.jpg

NOTE: In this metric, ‘automated export metrics’ is not used. This permits manual inspection and rejection of traces. The only measured parameter is peak current. Alternative measurements such as AUC or absolute current amplitude are not measured.

Cleanup Protocol

From the Cleanup Protocol menu, input the following criteria into each of the different folder tabs:

Image aea-Image040

NOTE: This protocol discards tips and sonicates the electrode plate in aqueous solution. To further minimize compound carryover, an additional manual wash of the e-plate in 70% ethanol was performed off the instrument.

Run Experiment

1.

On Set Default Folders window, select the proper default folders for where the data and protocol files will reside.

2.

Select the desired Setup, Channel, and Cleanup Protocols (as previously described) from the drop down menus.

3.

Load the instrument deck. Make sure tips, boats, and plates are seated correctly:

a.

Pipette tips are in position Tips (position labeled 1 in the diagram below)

b.

External buffer boat is in position External Buffer (position 2)

c.

Agonist plate is in position Compound 1 (position 3)

d.

Appropriate new patch plate (SH or PPC) is in position Patch Plate (position 4)

e.

Correct electrode plate is positioned on Wash Station (make sure that the plate is lying flat and in correct orientation) (position 5)

f.

Compound plate is in position Compound 2 (position 6)

g.

Clean cell boat is in position Cell Boat (position 7)

Image aea-Image041.jpg

4.

Make sure that the cells position is empty (conical vial has been removed).

5.

Set the experiment name.

Image aea-Image042.jpg

6.

Click the green Play button to start the experiment.

7.

When the instrument pauses to load cells (after Hole Test) add the conical vial with cell suspension.

8.

At the end of the run, remove and discard conical vial, tip box, agonist plate and compound plate. Remove the cell boat, wash it with H2O and wipe dry. Remove the electrode plate and wash by dipping in 70% ethanol followed by 0.9% NaCl solution. Buffer reservoir can be left on the deck if additional runs are planned.

Analyze Data

1.

From the Data Analysis button on the left, see below for an example of all the tabs that can be opened to configure data analysis and export.

Image aea-Image043.jpg

2.

In the Define Filters tab, set your desired hole/seal resistance, peak current amplitude, and/or other criteria for removing well data from the analysis. Additional filtering can be performed by visual inspection of traces and manual removal of wells.

3.

Click the Metric Exports button on the left and define how your metrics should be exported.

Image aea-Image044.jpg

End of Day Flush and Rinse

1.

Make certain 50% EtOH is in the alcohol wash bottle.

2.

Disconnect external solution bottle, amphotericin B bottle, and internal solution bottle and replace with empty 50 mL conical vials.

3.

Replace the electrode plate with model cell.

4.

Replace the contaminated patch plate with an un-contaminated (has not been exposed to compounds) used patch plate.

5.

Run End of Day Flush & Rinse protocol from the Utilities menu.

6.

Remove ground electrodes and place them back into internal solution soak.

7.

Power off the IonWorks Barracuda.

Modifications of IW-Barracuda Protocols for Assessment of Agonists and Allosteric Modulators of GABAA Channels

Agonist

Use the same protocol as described for assessment of current block. Compound plate should contain 1 mM GABA in position designated as maximum control (Col 1, 2, 23 and 24). In the analysis, select scan 1 instead of scan 2.

Positive Allosteric Modulator (PAM)

Use the same protocol as described for assessment of current block. Adjust the agonist concentration in “agonist Plate” to GABA EC20. Add appropriate positive and negative controls to “Agonist Plate” (Col. 1, 2, 23 and 24). Analyze scan 2.

Negative Allosteric Modulator (NAM)

Use the same protocol as described for assessment of current block. Adjust the agonist concentration in “agonist Plate” to GABA EC60. Add appropriate positive and negative controls to “Agonist Plate” (Col. 1, 2, 23 and 24). Analyze scan 2.

Appendix 4: Protocol for Assessment of State-Dependent Sodium Channel Inhibitors on PatchXpress

Overview

This protocol is designed to evaluate block of voltage-gated sodium channels at depolarizing voltages that favor interactions of test compounds with inactivated states of the channel.

Voltage Protocol using scripting

Cells are sealed and whole cell configuration attained according to defined patch protocols that can be obtained from Molecular Devices. Sampling rate is adjusted to the maximal available rate of 31.25 kHz (32 μs sampling interval). Once a stable whole cell configuration is achieved, cells are held at -120 mV and currents are monitored with short (5-20 ms) test pulses to a voltage near the peak of the current-voltage curve until currents reach stable amplitude. At this point, a standard protocol is executed to determine the voltage dependence of steady-state inactivation, for example this protocol may consist of a set of 1 s conditioning pulses to voltages ranging from -120 mV to -20 mV followed by a short test pulse to the same voltage used during the previous test pulses. A software script is used to select a new holding potential that corresponds to approximately 40% channel inactivation. Current amplitudes are again allowed to stabilize at the new inactivated potential for approximately 2-3 minutes while applying test pulses at 0.1 Hz. Following current stabilization at the inactivated potential, a 10 pulse train at 5 Hz is applied. Following the pulse train, currents are again allowed to stabilize at 0.1 Hz prior to the first compound addition. Two compound additions occur approximately 15 s apart, and effects on current amplitude are monitored by test pulses at 0.1 Hz. Once the software script determines that stable current amplitudes have been reached, a second 5 Hz pulse train determines use-dependent effects. Wash-out occurs at the inactivated potential for approximately one minute before the holding potential is returned to -120 mV with continued wash-out until current recovers to 80% of its initial control amplitude. Assuming satisfactory wash-out has occurred, the cell may be used for further compound additions until cell health and/or current amplitude do not remain stable.

Typical Solutions

External Buffer Solution: 137 mM NaCl, 4 mM KCl, 1.8 mM CaCl2, 0.2 mM MgCl2, 10 mM HEPES, 10 mM D-mannitol, adjust pH to 7.4 using NaOH. Final osmolality: 300-310 mOsm

Internal Buffer Solution: 60 mM CsF, 70 mM CsCl, 5 mM NaCl, 10 mM EGTA, 10 mM HEPES, adjust pH to 7.4 using CsOH. Final osmolality: 290-300 mOsm

Notes:

-Slightly lower osmolality in the internal solution compared to the external solution often helps with sealing.

-Some F- in the internal solution helps to maintain seal resistances; however, higher F- concentrations can lead to time-dependent shifts in the voltage dependence of inactivation.

-At room temperature the Na+ reversal potential for this solution is approximately +84 mV.

Start of Day

1.

Allow external and internal solutions to come to room temperature, if stored refrigerated.

2.

Turn on PatchXpress 7000A Instrument using switch located on left side of instrument.

3.

The associated CPU will need to be restarted to initialize settings between CPU and Hardware.

4.

Open the PatchXpress Control Software.

5.

When prompted by PatchXpress Commander, log on to the server housing the database.

6.

After initialization, the instrument will prompt you to begin Start of Day procedure.

7.

Make sure there is H2O in the wash (thin) container and that the waste (large) container inside the PatchXpress cabinet is empty.

8.

Load external and internal solutions onto the PatchXpress Instrument.

9.

Begin Start of Day procedure - this replaces the H2O in the lines with the internal and external assay buffers- using the blank seal chip stored in the instrument.

10.

Prime wash station (hardware fluidics).

11.

Add pipette tips.

Cell preparation

1.

Start with >70% confluent T-75 flask (or equivalent) of HEK cells expressing the Nav1 channel of interest.

2.

Pour off media, and add 2 mL TrypLE .

3.

Incubate with TrypLE for 30-120s, tap flask to dislodge cells and add 3 mL of media.

4.

Triturate with 5 mL pipette.

5.

Transfer to 15 mL conical tube and spin at 1300-1500 rpm for 3 min.

6.

Aspirate supernatant and resuspend cell pellet in 4-8 mL of media.

7.

Triturate with 1 mL pipette and add 1 mL each of the cell suspension to 4-6 small Eppendorf tubes.

8.

Store uncapped in incubator for at least 30 min and up to 3 h (optimal timing may differ between cell lines).

9.

At the start of an experiment, spin cells at 2000 rpm for 2 min, remove supernatant, resuspend in 200 μL external buffer solution, triturate lightly and place the tube into the PatchXpress instrument

Start of Experiment

The following settings must be established to define the experiment on the PatchXpress prior to loading the Seal chip:

1.

In Define Experiment window, select the proper project from the menu.

2.

Create a table using Excel that specifies the location and concentration of each compound to be tested in the compound plate and select this table in the PatchXpress software.

Image aea-Image045.jpg

3.

Select the proper procedure from the dropdown box in Procedure window:

Image aea-Image046.jpg

4.

Check Re-use cells with 10 replicates.

5.

Under Select Cell, select cell, i.e. the parental cell background – typically HEK293 or CHO, and channel, e.g. Nav1.x, from the drop down boxes.

6.

Select the appropriate internal and external solutions.

7.

Select the appropriate patch settings. Molecular Devices provides standard settings for HEK293 and CHO cells that can be modified for optimal results with a particular cell line of interest.

8.

Select Screening mode which means that the software will keep track of which compounds and concentrations have been tested successfully.

Once the experiment has been defined:

  • Place test compounds onto the deck of the PatchXpress instrument. Test compounds need to be dissolved in external buffer solution and can be presented in a 96-well plate or in glass vials inserted into a plastic rack provided by Molecular Devices.
  • Wash station may need to be primed again, if too much time has passed since this step was performed last.
  • Resuspend an aliquot of cells in External Buffer Solution (see last step of Cell Preparation protocol) and place into the PatchXpress instrument.
  • Remove a seal chip from the glass vial and flick out the remaining water. Blow the seal chip dry using the tool inside the PatchXpress cabinet and place the seal chip into the receiver on the deck of the PatchXpress with the straight edge to the back.
  • Click the green Play button to start the experiment. Once the instrument initializes, it will request a barcode. Enter the barcode from the seal chip vial using the barcode reader (this step is optional).
  • After the seal chip is loaded, the instrument will “call for cells”, requiring the user to enter information regarding the cell line for future reference (examples include cell line, passage number and time after dissociation).

Analysis

Aggregate data from multiple single concentration additions are plotted and fit using the Hill equation. Use-dependence based on the last pulse of the 5 Hz pulse trains may also be reported. Run down correction may be employed on a limited basis.

This procedure overview is summarized below for a compound which demonstrates inhibition:

Image aea-Image047

End of Day

1.

Replace external and internal buffer solutions with water.

2.

Run End of Day protocol.

3.

Turn off the computer and PatchXpress instrument.

Copyright Notice

All Assay Guidance Manual content, except where otherwise noted, is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported license (CC BY-NC-SA 3.0), which permits copying, distribution, transmission, and adaptation of the work, provided the original work is properly cited and not used for commercial purposes. Any altered, transformed, or adapted form of the work may only be distributed under the same or similar license to this one.

Bookshelf ID: NBK424997PMID: 28277626

Views

Assay Guidance Manual Links

Related information

  • PMC
    PubMed Central citations
  • PubMed
    Links to PubMed

Similar articles in PubMed

See reviews...See all...

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...