Format

Send to

Choose Destination
Magn Reson Med. 2002 Oct;48(4):707-14.

Asymmetrical gradient coil for head imaging.

Author information

1
Medical Department, Brookhaven National Laboratory, Upton, New York 11973, USA. tomasi@bnl.gov

Abstract

This work presents a novel approach to develop dedicated transverse gradient coils for head imaging. The proposed coil design is based on the stochastic optimization of an asymmetrical stream function and improves the matching between the region-of-interest and the homogeneous gradient volume. Additionally, the electric field produced by these asymmetrical coils is 30% lower than that produced by standard symmetrical designs, which minimizes the risk of magnetostimulation of nerves in fast imaging techniques. A prototype of the asymmetrical gradient coil was built to test the method and magnetic field produced by the prototype was measured. Magnetic field measurements and electrical parameters of coils are in good agreement with theoretical calculations.

PMID:
12353289
DOI:
10.1002/mrm.10263
[Indexed for MEDLINE]
Free full text

Supplemental Content

Full text links

Icon for Wiley
Loading ...
Support Center