  • Showing results for rapid measurement of human milk macronutrient in the neonatal intensive care unit accuracy and precision of fourier transform mid-infrared spectroscopy. Your search for apid measurement of human milk macronutrients in the neonatal intensive care unit: accuracy and precision of fourier transform mid-infrared spectroscopy retrieved no results.
J Hum Lact. 2014 May;30(2):180-9. doi: 10.1177/0890334413517941. Epub 2014 Jan 14.

Rapid measurement of human milk macronutrients in the neonatal intensive care unit: accuracy and precision of fourier transform mid-infrared spectroscopy.

Author information

1
1Department of Food Science and Technology, University of California Davis, Davis, CA, USA.

Abstract

BACKGROUND:

Although it is well established that human milk varies widely in macronutrient content, it remains common for human milk fortification for premature infants to be based on historic mean values. As a result, those caring for premature infants often underestimate protein intake. Rapid precise measurement of human milk protein, fat, and lactose to allow individualized fortification has been proposed for decades but remains elusive due to technical challenges.

OBJECTIVE:

This study aimed to evaluate the accuracy and precision of a Fourier transform (FT) mid-infrared (IR) spectroscope in the neonatal intensive care unit to measure human milk fat, total protein, lactose, and calculated energy compared with standard chemical analyses.

METHODS:

One hundred sixteen breast milk samples across lactation stages from women who delivered at term (n = 69) and preterm (n = 5) were analyzed with the FT mid-IR spectroscope and with standard chemical methods. Ten of the samples were tested in replicate using the FT mid-IR spectroscope to determine repeatability.

RESULTS:

The agreement between the FT mid-IR spectroscope analysis and reference methods was high for protein and fat and moderate for lactose and energy. The intra-assay coefficients of variation for all outcomes were less than 3%.

CONCLUSION:

The FT mid-IR spectroscope demonstrated high accuracy in measurement of total protein and fat of preterm and term milk with high precision.

KEYWORDS:

Fourier transform; breast milk; breastfeeding; fat; lactose; mid-infrared spectroscopy; nutrition; point-of-care testing; preterm infants; protein

PMID:
24423984
DOI:
10.1177/0890334413517941
[Indexed for MEDLINE]