Determination of starane (fluroxypyr) herbicide using flow injection spectrophotometry

Anal Sci. 2006 Jan;22(1):145-6. doi: 10.2116/analsci.22.145.

Abstract

The starane herbicide was spectrophotometrically determined by the diazotization method in a flow injection assembly. Since starane is a substituted pyridyl compound the NH2 group at the p-position was exploited for diazotization. Starane was diazotized with nitrite and the diazotized product is coupled with beta-naphthol. The absorbance of the resulting azo dye was measured at 395 nm with a molar absorptivity of 1.5 x 10(4) L mol(-1) cm(-1). The calibration graph was linear over the range of 0.6 to 10 microg/mL, with a relative standard deviation (RSD) of 1.67% and a sampling through put of 60 samples h(-1). The % recovery for the determination of starane was found to be 96%. The method was successfully applied to the determination of the active ingredient of starane herbicide in its formulation as well as in food samples.

MeSH terms

  • Acetates / analysis*
  • Calibration
  • Flow Injection Analysis / instrumentation
  • Flow Injection Analysis / methods
  • Herbicides / analysis*
  • Molecular Structure
  • Pyridines / analysis*
  • Sensitivity and Specificity
  • Spectrophotometry / instrumentation
  • Spectrophotometry / methods

Substances

  • Acetates
  • Herbicides
  • Pyridines
  • 4,5-amino-3,5-dichloro-6-fluoro-2-pyridinyloxyacetic acid