Format

Send to

Choose Destination
See comment in PubMed Commons below
Schweiz Arch Tierheilkd. 1998;140(12):507-12.

[Evaluation of the portable blood analyser i-STAT].

[Article in German]

Author information

1
Klinik für Nutztiere und Pferde der Universität Bern.

Abstract

The purpose of this study was to compare the results of horse blood examinations on a portable blood analysis system, i-STAT SDI, with conventional laboratory equipment and to verify the accuracy and precision. The precision within run was high and the coefficients of variation ranged from 0 to 3.85%. The comparative analysis of patient samples with routine methods showed a very strong positive correlation with correlation coefficients above 0.96 for blood gases, pH, glucose, urea, hemoglobin, hematocrit, sodium and potassium, with the exception of chloride where it was 0.867. The i-STAT system provides a quick and accurate stable-side evaluation of laboratory data in critically ill horses and emergency situations.

PMID:
9863357
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Loading ...
    Support Center