J Chromatogr B Biomed Sci Appl. 1998 Apr 24;708(1-2):55-60.

Analysis of cysteine and N-acetylcysteine in human plasma by high-performance liquid chromatography at the basal state and after oral administration of N-acetylcysteine.

Author information

1
Institute of Clinical Pharmacology, Hannover Medical School, Germany.

Abstract

A high-performance liquid chromatographic method for the determination of free reduced cysteine and N-acetylcysteine in human plasma at the basal state and after oral administration of N-acetylcysteine is described. The method is based on acid-catalysed conversion of plasma thiols to the corresponding S-nitroso derivatives by excess of nitrite and their subsequent cation-pairing RP-HPLC with detection at 333 nm. Recovery rates of cysteine and N-acetylcysteine added to human plasma were 94.6 and 99.6%, respectively. Inter- and intra-day precision were below 6%. In healthy humans (n = 5), free reduced cysteine was determined to be (mean+/-S.E.) 10.0+/-0.96 microM. No N-acetylcysteine was detected in plasma of these subjects above the limit of detection (e.g. 170 nM). The method was successfully applied to a pharmacokinetic study on orally administered N-acetylcysteine to healthy volunteers.

PMID:
9653946
DOI:
10.1016/s0378-4347(97)00670-1
[Indexed for MEDLINE]

Supplemental Content

Loading ...