Anaesth Intensive Care. 1993 Apr;21(2):204-7.

Laboratory evaluation of the Go Medical PCI: a disposable patient-controlled analgesia device.

Author information

1
Department of Anaesthesia and Intensive Care, Flinders Medical Centre, Bedford Park, South Australia.

Abstract

A low-cost, disposable patient-controlled analgesia device was tested using a computer-controlled, automated testing station designed to simulate clinical use. Five devices were tested and delivered a mean bolus dose size of 95% of nominal, although one device achieved only 85% of nominal. Delivery was lowest immediately following an eight-hour interval during which no demands were made. Further development and evaluation of this interesting device is warranted.