Ultramicroscopy. 1993 Mar;48(3):347-58.

Use of radial density plots to calibrate image magnification for frozen-hydrated specimens.

Author information

1
Department of Biological Sciences, Purdue University, West Lafayette, IN 47907.

Abstract

Accurate magnification calibration for transmission electron microscopy is best achieved with the use of appropriate standards and an objective calibration technique. We have developed a reliable method for calibrating the magnification of images from frozen-hydrated specimens. Invariant features in radial density plots of a standard are compared with the corresponding features in a "defocused" X-ray model of the same standard. Defocused X-ray models were generated to mimic the conditions of cryo-electron microscopy. The technique is demonstrated with polyoma virus, which was used as an internal standard to calibrate micrographs of bovine papilloma virus type 1 and bacteriophage phi X174. Calibrations of the micrographs were estimated to be accurate to 0.35%-0.5%. Accurate scaling of a three-dimensional structure allows additional calibrations to be made with radial density plots computed from two- or three-dimensional data.