Format

Send to

Choose Destination
Am J Otol. 1993 Jul;14(4):386-91.

Radiologic evaluation of multichannel intracochlear implant insertion depth.

Author information

1
Department of Otolaryngology, Holt-Krock Clinic, Fort Smith, Arkansas 72901.

Erratum in

  • Am J Otol 1993 Nov;14(6):627.

Abstract

Postoperative plain film x-rays are necessary in all multichannel cochlear implant patients to confirm intracochlear position, detect possible electrode kinking, and provide a reference if postoperative slippage occurs. In addition, precise documentation of multichannel intracochlear electrode insertion depths is necessary for comparison of speech recognition results among patients and may be of use for future speech processing strategies. In the present study, a method has been devised, using a modified Stenver's view, to more accurately document insertion depths of the electrode array and location of individual electrodes on 50 multichannel cochlear implant patients. Surgical estimates of insertion depth are shown to have great variability in regard to distance along the basilar membrane when compared with x-ray documentation. Additionally, there is preliminary evidence that insertion depth, as determined by x-ray studies, has a strong correlation with open-set speech discrimination.

PMID:
8238277
[Indexed for MEDLINE]

Supplemental Content

Loading ...
Support Center