Migration of corrosion products from modular hip prostheses. Particle microanalysis and histopathological findings.

Author information

1
Department of Orthopedic Surgery, Rush-Presbyterian-St. Luke's Medical Center, Chicago, Illinois 60612.

Abstract

Migration of solid corrosion products from the modular head-neck junction of fifteen total hip replacements to the periprosthetic tissues was studied. The devices and tissues were recovered at the time of a revision procedure or at autopsy after a mean of sixty-four months (range, eight to ninety-seven months). The prostheses had a cobalt-chromium-alloy head coupled with a cobalt-chromium-alloy or a titanium-alloy stem. The solid corrosion product was identified by electron microprobe analysis and Fourier transform infrared microprobe spectroscopy as a chromium orthophosphate hydrate-rich material. The product was present at the junction of the modular head and neck and as particles within the periprosthetic tissues as early as eight months postoperatively. In several hips, it was also present on the polyethylene bearing surface. The particles in the tissues ranged in size from less than one to 500 micrometers. They were present within histiocytes or were surrounded by foreign-body giant cells in the pseudocapsule of the hip joint; in the membranes of the femoral bone-implant interface; and at sites of femoral endosteal erosions, with and without loosening of the femoral component.

PMID:
8077264