Acta Anaesthesiol Scand. 1995 Feb;39(2):268-9.

Prolonged interference of blue dye "patent blue" with pulse oximetry readings.

Author information

1
Department of Anesthesiology and Reanimatology, Gunma University School of Medicine, Maebashi-shi, Japan.

Abstract

Several blue dyes were reported to interfere with the accurate determination of oxygen saturation by pulse oximetry. However, in previous reports the interferences were observed only for a short duration. We have experienced a case in which the interference by a blue dye, "patent blue" continued for a long time. It was suggested that the blue dye, when it was injected intra-arterially and in an anaemic patient, interfered with the pulse oximetry measurement to a greater extent.