Quantitative mammography contrast threshold test tool

Med Phys. 1995 Feb;22(2):127-32. doi: 10.1118/1.597462.

Abstract

Mammographic contrast is commonly evaluated by visualizing small objects of varying size or mass divided by projected area. These qualitative contrast determinations are commonly performed by imaging a phantom like the American College of Radiology accreditation phantom at clinical mammographic settings. However, this contrast assessment does not take into account the kVp of the machine. This work describes a quantitative mammography contrast threshold test tool which examines light object contrast on a uniform background for a contrast range of 0.32% to 1.38% at 25 kVp. For this mammography contrast threshold test tool, contrast is defined by delta I/I = loge (psi O/ psi b), where psi O is the target energy flux, and psi b is the background energy flux. Contrast threshold is defined as the lowest contrast value for which the objects are visible. Unlike traditional assessments of mammographic contrast, this measurement of contrast threshold is kVp corrected. The mammography contrast threshold test tool is constructed out of common plastics and provides a quantitative means of assessing contrast threshold for individual mammographic units and total mammographic systems.

Publication types

  • Comparative Study

MeSH terms

  • Accreditation
  • Contrast Media
  • Female
  • Humans
  • Mammography / methods*
  • Mammography / standards
  • Mathematics
  • Phantoms, Imaging*
  • Sensitivity and Specificity
  • Societies, Medical
  • United States

Substances

  • Contrast Media