Low-frequency response characteristics of three Grass model 7 polygraphs

J Appl Physiol (1985). 1985 Mar;58(3):1026-30. doi: 10.1152/jappl.1985.58.3.1026.

Abstract

A low-frequency response analysis of three Grass model 7 polygraphs was undertaken. Observed error was generally found to fall within the manufacturer's stated range of +5 to -10% of DC signal height over the frequency range of human respiration (0.1-3 Hz), but this was not the case for frequencies greater than 6 Hz under certain circumstances. The magnitude of error was seen to vary directly with frequency and indirectly with pen-deflection amplitude and paper speed. The pen-oscillograph apparatus was the predominant source of low-frequency error, and this is probably due to pen inertia and pen friction on the writing surface. Two schemes to reduce such error are presented.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Calibration
  • Equipment Failure
  • Humans
  • Oscillometry
  • Physiology / instrumentation*
  • Respiration*