Teleglaucoma Using a New Smartphone-Based Tool for Visual Field Assessment

J Glaucoma. 2023 Mar 1;32(3):186-194. doi: 10.1097/IJG.0000000000002153. Epub 2022 Nov 29.

Abstract

Prcis: COVID-19 underlines the importance of telemedical diagnostics. The smartphone-based campimetry (Sb-C) is a newly developed digital application allowing visual field testing using a head-mounted device and a smartphone. It enables visual field screening remotely from a clinic.

Background: Sb-C is a newly developed tool for functional ophthalmic diagnosis. This study aimed to examine the comparability of the Sb-C and Octopus 900 to ensure ophthalmological care in times of social distancing.

Methods: Total 93 eyes were included in the study. After an ophthalmological examination, the visual field was tested by the Octopus program G1 and by the smartphone-based campimeter. The Sb-C was performed using VR glasses and an iPhone 6. The software Sb-C was downloaded and installed as SmartCampiTracker app and is examining the 30-degree visual field with 59 test positions corresponding to the G pattern of Octopus G1. Sensitivities were recorded and saved on the app. In addition, test-retest reliability was tested on 6 ophthalmologically healthy participants.

Results: The group comprised 48 women and 45 men (mean age: 62.52±12.2 y) including 19 controls, 17 patients with ocular hypertension, 11 preperimetric glaucomas, and 46 perimetric glaucomas. The mean sensitivity (MS) of all points of G1 perimetry was 23.13 dB (95% CI, 22.08-24.18). The MS of the Sb-C was 21.23 dB (95% CI, 20.37-22.08). The correlation between the mean MS measured by G1 perimetry and the Sb-C was strong ( r =0.815, P <0.05). The test-retest reliability showed a correlation of r =0.591 ( P <0.05) .

Conclusions: With some technical adjustments, the Sb-C shows promise for screening glaucoma and monitoring disease progression remotely from an ophthalmologic clinic.