Consistency between Impedance Technique and Echocardiogram Hemodynamic Measurements in Neonates

Am J Perinatol. 2021 Oct;38(12):1259-1262. doi: 10.1055/s-0040-1710030. Epub 2020 Apr 25.

Abstract

Objective: The aim of this study was to validate impedance technique (IT) by investigating the agreement in cardiac output measurements performed by IT and echocardiography (ECHO).

Study design: This is a prospective observational study, including a total of 30 neonates who underwent hemodynamic measurements by IT and ECHO. To determine the agreement between both methods, we performed IT to measure stroke volume (SV-IT) and cardiac output (CO-IT) immediately before or after ECHO to measure SV (SV-ECHO) and CO (CO-ECHO). The precision and accuracy of the IT relative to ECHO were assessed.

Results: SV-ECHO and SV-IT were (4.45 ± 0.78) and (4.54 ± 0.81) mL, respectively. The bias and limits of agreement of SV-IT were 0.09 mL and ( -1.92 to 1.73) mL, respectively. The true precision of SV-IT was 27.3%. Furthermore, CO-ECHO and CO-IT were (0.62 ± 0.12) and (0.61 ± 0.12) L/min, respectively. The bias and LoA of CO-IT were 0.01L/min and (-0.33 to 0.31) L/min, respectively. The true precision of CO-IT was 28.3%.

Conclusion: Agreement between the IT and ECHO in the cardiac output measurement appeared acceptable. However, the accuracy and precision of the IT approach should be further investigated using a larger sample.

Publication types

  • Observational Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Cardiac Output*
  • Echocardiography*
  • Electric Impedance*
  • Female
  • Humans
  • Infant, Newborn / physiology*
  • Infant, Premature / physiology
  • Male
  • Monitoring, Physiologic / instrumentation
  • Monitoring, Physiologic / methods*
  • Prospective Studies