Background: Total hip replacement is recognised as a major risk factor for deep vein thrombosis (DVT). The aim of this study was to investigate the feasibility of using a novel neuromuscular electrical stimulation device (NMES) for DVT prevention in patients recovering from elective hip replacement surgery.
Methods: Twenty-eight patients undergoing total hip replacement were randomised to receive postoperative treatment with either the NMES device or compression stockings continually from post-surgery until discharge (day 4). The primary outcome measure was the presence of symptomatic or asymptomatic DVT at 48 hours post-surgery and on the day of discharge from hip replacement surgery, as assessed by Duplex ultrasound. Secondary outcomes included hemodynamic responses to the devices, lower limb oedema, sit-to-stand and timed-up-and-go (TUG) scores, and hip range of motion.
Results: In the compression stockings group, two cases of asymptomatic DVT were identified by Duplex ultrasound at 48 hours post-surgery. No cases were found in the NMES group. Patients in the NMES group demonstrated a general trend of a decrease in leg volume from post-surgery to discharge, whereas leg volume largely remained static for the compression stockings group. In addition, positive hemodynamic effects were found in favour of the NMES group in the non-operated leg. The change in TUG scores also favoured the NMES group (NMES: 150 ± 152%, compression stockings: 363 ± 257% (p=0.03)), whereas no differences in sit-to-stand scores or hip range of motion were observed.
Conclusions: This study supports the feasibility of NMES as an alternative mechanical prophylaxis worn in the postoperative phase until discharge and provides important findings for clinicians considering novel mechanical prophylaxis options.