Optimization of acquisition parameters for reduced-dose thoracic CT: A phantom study

Diagn Interv Imaging. 2020 May;101(5):269-279. doi: 10.1016/j.diii.2020.01.012. Epub 2020 Feb 24.

Abstract

Purpose: The purpose of this study was to analyze the impact of different options for reduced-dose computed tomography (CT) on image noise and visibility of pulmonary structures in order to define the best choice of parameters when performing ultra-low dose acquisitions of the chest in clinical routine.

Materials and methods: Using an anthropomorphic chest phantom, CT images were acquired at four defined low dose levels (computed tomography dose index [CTDIvol]=0.15, 0.20, 0.30 and 0.40mGy), by changing tube voltage, pitch factor, or rotation time and adapting tube current to reach the predefined CTDIvol-values. Images were reconstructed using two different levels of iteration (adaptive statistical iterative reconstruction [ASIR®]-v70% and ASIR®-v100%). Signal-to-noise ratio (SNR) as well as contrast-to-noise ratio (CNR) was calculated. Visibility of pulmonary structures (bronchi/vessels) were assessed by two readers on a 5-point-Likert scale.

Results: Best visual image assessments and CNR/SNR were obtained with high tube voltage, while lowest scores were reached with lower pitch factor followed by high tube current. Protocols favoring lower pitch factor resulted in decreased visibility of bronchi/vessels, especially in the periphery. Decreasing radiation dose from 0.40 to 0.30mGy was not associated with a significant decrease in visual scores (P<0.05), however decreasing radiation dose from 0.30mGy to 0.15mGy was associated with a lower visibility of most of the evaluated structures (P<0.001). While image noise could be significantly reduced when ASIR®-v100% instead of ASIR®-v70% was used, the visibility-scores of pulmonary structures did not change significantly.

Conclusion: Favoring high tube voltage is the best option for reduced-dose protocols. A decrease of SNR and CNR does not necessarily go along with reduced visibility of pulmonary structures.

Keywords: Phantoms; Radiation dosage; Thorax; Tomography; X-ray computed; imaging.

MeSH terms

  • Algorithms
  • Humans
  • Phantoms, Imaging
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted*
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed*