In vitro study of trileaflet polytetrafluoroethylene conduit and its valve-in-valve transformation

Interact Cardiovasc Thorac Surg. 2020 Mar 1;30(3):408-416. doi: 10.1093/icvts/ivz274.

Abstract

Objectives: Handmade trileaflet expanded polytetrafluoroethylene valved conduit developed using the flip-over method has been tailored for pulmonary valve reconstruction with satisfactory outcomes. We investigated the in vitro performance of the valve design in a mock circulatory system with various conduit sizes. In our study, the design was transformed into a transcatheter stent graft system which could fit in original valved conduits in a valve-in-valve fashion.

Methods: Five different sizes of valved polytetrafluoroethylene vascular grafts (16, 18, 20, 22 and 24 mm) were mounted onto a mock circulatory system with a prism window for direct leaflets motion observation. Transvalvular pressure gradients were recorded using pressure transducers. Mean and instant flows were determined via a rotameter and a flowmeter. Similar flip-over trileaflet valve design was then carried out in 3 available stent graft sizes (23, 26 and 28.5 mm, Gore aortic extender), which were deployed inside the valved conduits.

Results: Peak pressure gradient across 5 different sized graft valves, in their appropriate flow setting (2.0, 2.5 and 5.0 l/min), ranged from 4.7 to 13.2 mmHg. No significant valve regurgitation was noted (regurgitant fraction: 1.6-4.9%) in all valve sizes and combinations. Three sizes of the trileaflet-valved stent grafts were implanted in the 4 sizes of valved conduits except for the 16-mm conduit. Peak pressure gradient increase after valved-stent graft-in-valved-conduit setting was <10 mmHg in all 4 conduits.

Conclusions: The study showed excellent in vitro performance of trileaflet polytetrafluoroethylene valved conduits. Its valved stent graft transformation provided data which may serve as a reference for transcatheter valve-in-valve research in the future.

Keywords: Extended polytetrafluoroethylene valve; Handmade trileaflet expanded polytetrafluoroethylene valved conduit; Pulmonary valve replacement; Trileaflet expanded polytetrafluoroethylene valved stent graft; Valve-in-valve.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Vessel Prosthesis*
  • Heart Valve Diseases / surgery
  • Heart Valve Prosthesis Implantation
  • Heart Valve Prosthesis*
  • Humans
  • Materials Testing
  • Models, Cardiovascular
  • Polytetrafluoroethylene*
  • Prosthesis Design*
  • Pulmonary Valve*
  • Stents*

Substances

  • Polytetrafluoroethylene