Has the threshold for revision surgery for adverse reactions to metal debris changed in metal-on-metal hip arthroplasty patients? A cohort study of 239 patients using an adapted risk-stratification algorithm

Acta Orthop. 2019 Dec;90(6):530-536. doi: 10.1080/17453674.2019.1659661. Epub 2019 Sep 9.

Abstract

Background and purpose - A risk-stratification algorithm for metal-on-metal hip arthroplasty (MoMHA) patients was devised by US experts to help clinicians make management decisions. However, the proposed algorithm did not cover all potential patient or surgical abnormalities. Therefore we adapted the US risk-stratification algorithm in MoMHA patients revised for adverse reactions to metal debris (ARMD) to determine the variability in the revision threshold, and also whether high-risk patients had inferior outcomes following revision.Patients and methods - We analysed 239 MoMHA revisions for ARMD between 2001 and 2016 from 2 centres with pre-revision blood metal ions and imaging. Patients were stratified (low risk, moderate risk, high risk) using pre-revision factors (implant, radiographic, blood metal ions, cross-sectional imaging) by adapting a published algorithm. The risk categories for each factor were assessed against revision year, revision centre, and post-revision outcomes (re-revision surgery, and any poor outcome).Results - Compared with hips revised before 2012, hips revised from 2012 onwards included more high-risk implants (44% vs. 17% pre-2012), high-risk radiographic features (85% vs. 69% pre-2012), and low-risk metal ions (41% vs. 19% pre-2012). 1 centre more frequently revised patients with high-risk implants (48% vs. 14%) and low-risk blood metal ions (45% vs. 15%) compared with the other. All these comparisons were statistically significant (p < 0.05). With the limited sample size available, implant, radiographic, blood metal ion, and cross-sectional imaging risk groups did not statistically significantly affect the rates of re-revision surgery or frequency of poor outcomes post-revision.Interpretation - When applying the adapted risk-stratification algorithm the threshold for ARMD revision changed over time, presumably due to increasing evidence, patient surveillance, and investigation since 2012. Lower blood metal ion thresholds were used from 2012 for ARMD revisions; however, there was evidence that centres attached different importance to metal ions when managing patients. High-risk patients did not have inferior outcomes following ARMD revision.

MeSH terms

  • Algorithms
  • Arthroplasty, Replacement, Hip*
  • Chromium / blood
  • Cobalt / blood
  • Cohort Studies
  • Female
  • Foreign-Body Reaction / etiology
  • Foreign-Body Reaction / surgery
  • Humans
  • Ions / blood
  • Male
  • Metal-on-Metal Joint Prostheses / adverse effects*
  • Middle Aged
  • Reoperation*
  • Retrospective Studies
  • Risk Assessment*

Substances

  • Ions
  • Chromium
  • Cobalt