Med Phys. 2019 Aug 25. doi: 10.1002/mp.13785. [Epub ahead of print]

NEMA NU-4 2008 Performance Evaluation of Xtrim-PET: A prototype SiPM-based preclinical scanner.

Author information

1
Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran.
2
Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran.
3
Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran.
4
PET/CT and Cyclotron Center, Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
5
Research Center for Nuclear Medicine, Shariati Hospital, Tehran University of Medical Sciences, Tehran, IRAN.
6
Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, Geneva, CH-1211, Switzerland.
7
Geneva University Neurocenter, Geneva University, CH-1205, Geneva, Switzerland.
8
Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, 9700 RB, Groningen, Netherlands.
9
Department of Nuclear Medicine, University of Southern Denmark, 500, Odense, Denmark.

Abstract

PURPOSE:

Xtrim-PET is a newly designed Silicon Photomultipliers (SiPMs)-based prototype PET scanner dedicated for small laboratory animal imaging. We present the performance evaluation of the Xtrim-PET scanner following NEMA NU-4 2008 standards to help optimizing scanning protocols which can be achieved through standard and reliable system performance characterization.

METHODS:

The performance assessment was conducted according to the National Electrical Manufacturers Association (NEMA) NU-4 2008 standards in terms of spatial resolution, sensitivity, counting rate performance, scatter fraction and image quality. The in vivo imaging capability of the scanner is also showcased through scanning a normal mouse injected with 18 F-FDG. Furthermore, the performance characteristics of the developed scanner are compared with commercially available systems and current prototypes.

RESULTS:

The volumetric spatial resolution at 5 mm radial offset from the central axis of the scanner is 6.81 μl whereas a peak absolute sensitivity of 2.99% was achieved using a 250-650 keV energy window and a 10 ns timing window. The peak noise-equivalent count rate (NECR) using a mouse-like phantom is 113.18 kcps at 0.34 KBq/cc with 12.5% scatter fraction, whereas the NECR peaked at 82.76 kcps for an activity concentration level of 0.048 KBq/cc with a scatter fraction of 25.8% for rat-like phantom. An excellent uniformity (3.8%) was obtained using NEMA image quality phantom. Recovery coefficients of 90%, 86%, 68%, 40% and 12% were calculated for rod diameters of 5, 4, 3, 2 and 1 mm, respectively. Spill-over ratios for air-filled and water-filled chambers were 35% and 25% without applying any correction for attenuation and Compton scattering effects.

CONCLUSION:

Our findings revealed that beyond compactness, lightweight, easy installation and good energy resolution, the Xtrim-PET prototype presents a reasonable performance making it suitable for preclinical molecular imaging-based research. This article is protected by copyright. All rights reserved.

KEYWORDS:

PET ; NEMA standard; detectors; performance evaluation; small animal imaging