Format

Send to

Choose Destination
Indian J Nucl Med. 2019 Jul-Sep;34(3):201-204. doi: 10.4103/ijnm.IJNM_42_19.

Comparison of Image Quality of Different Radionuclides Technetium-99m, Samarium-153, and Iodine-123.

Author information

1
Department of Physics, LPHE, Modeling and Simulations, Faculty of Science, Mohammed V University, Rabat, Morocco.
2
Department of Medical Physics, Academic Hospital of Udine, Udine, Italy.

Abstract

Introduction:

The choice of the radionuclide has a key role in nuclear medicine which appearing the lowest scatter fraction. In addition, the presence of penetrated and scattered photons from collimator in single-photon emission computed tomography images degrades resolution and contrast. Thus, image quality depends on sensitivity and resolution of the collimator-detector system. The goal of this study was to compare the image quality that can be achieved by three radionuclides: technetium-99 m (Tc-99 m), iodine-123 (I-123), and samarium-153 (Sm-153).

Materials and Methods:

Tc-99 m and Sm-153 were imaged with low-energy high resolution (LEHR) collimator, while I-123 was imaged with medium-energy (ME) collimator. We modeled the Siemens Symbia Medical system using Monte Carlo simulation code SIMIND. The imaging characteristics of each radionuclide were investigated by simulated data: point spread function, sensitivity (Cps/MBq) and geometric, penetration and scattering distribution.

Results:

Tc-99 m and Sm-153 give best and results with LEHR collimator for spatial resolution (full width at half maximum [FWHM] = 3.19 mm; full width at tenth maximum [FWTM] = 6.73 mm) and (FWHM = 3.22 mm; FWTM = 7.39 mm), respectively. Whereas, I-123 provided with ME collimator a lower resolution (FWHM = 4.89 mm; FWTM = 9.89 mm). The sensitivity recorded by Tc-99 m, Sm-153, and I-153 were (31.21 Cps/MBq), (10.16 Cps/MBq), and (51.22 Cps/MBq), respectively.

Conclusion:

Tc-99 m and Sm-153 give the best and generally similar imaging properties with LEHR. For I-123, the ME collimator helps lowering the influence of high-energy gamma rays.

KEYWORDS:

Geometric; SIMIND; imaging; penetration; resolution; scattering; sensitivity