The evaluation of a hybrid biomechanical deformable registration method on a multistage physical phantom with reproducible deformation

Radiat Oncol. 2018 Dec 4;13(1):240. doi: 10.1186/s13014-018-1192-x.

Abstract

Background: Advanced clinical applications, such as dose accumulation and adaptive radiation therapy, require deformable image registration (DIR) algorithms capable of voxel-wise accurate mapping of treatment dose or functional imaging. By utilizing a multistage deformable phantom, the authors investigated scenarios where biomechanical refinement method (BM-DIR) may be better than the pure image intensity based deformable registration (IM-DIR).

Methods: The authors developed a biomechanical-model based DIR refinement method (BM-DIR) to refine the deformable vector field (DVF) from any initial intensity-based DIR (IM-DIR). The BM-DIR method was quantitatively evaluated on a novel phantom capable of ten reproducible gradually-increasing deformation stages using the urethra tube as a surrogate. The internal DIR accuracy was inspected in term of the Dice similarity coefficient (DSC), Hausdorff and mean surface distance as defined in of the urethra structure inside the phantom and compared with that of the initial IM-DIR under various stages of deformation. Voxel-wise deformation vector discrepancy and Jacobian regularity were also inspected to evaluate the output DVFs. In addition to phantom, two pairs of Head&Neck patient MR images with expert-defined landmarks inside parotids were utilized to evaluate the BM-DIR accuracy with target registration error (TRE).

Results: The DSC and surface distance measures of the inner urethra tube indicated the BM-DIR method can improve the internal DVF accuracy on masked MR images for the phases of a large degree of deformation. The smoother Jacobian distribution from the BM-DIR suggests more physically-plausible internal deformation. For H&N cancer patients, the BM-DIR improved the TRE from 0.339 cm to 0.210 cm for the landmarks inside parotid on the masked MR images.

Conclusions: We have quantitatively demonstrated on a multi-stage physical phantom and limited patient data that the proposed BM-DIR can improve the accuracy inside solid organs with large deformation where distinctive image features are absent.

Keywords: Biomechanical model; Deformable image registration; IGRT; adaptive radiotherapy; dose accumulation.

MeSH terms

  • Algorithms*
  • Head and Neck Neoplasms / diagnostic imaging
  • Head and Neck Neoplasms / pathology
  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Male
  • Phantoms, Imaging*
  • Prostate / diagnostic imaging
  • Prostate / pathology
  • Prostate / radiation effects*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Intensity-Modulated / methods
  • Tomography, X-Ray Computed / methods
  • Urethra / diagnostic imaging
  • Urethra / pathology
  • Urethra / radiation effects*