Blood Press Monit. 2018 Dec;23(6):312-314. doi: 10.1097/MBP.0000000000000347.

Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2: 2013 standard.

Author information

1
Department of Medicine, Division of Cardiovascular Medicine, Jichi Medical University School of Medicine, Tochigi, Japan.

Abstract

OBJECTIVE:

The aim of this study was to validate the UA-1200BLE device using the ISO 81060-2:2013 standard.

MATERIALS AND METHODS:

Individuals aged more than 12 years were recruited from among the outpatients and volunteers of Jichi Medical University (Tochigi, Japan). Eighty-five patients were included in the final analyses. Validation was performed using the same-arm sequence protocol as per the ISO 81060-2:2013 standard.

RESULTS:

The device performed well against the standard; the mean and SD values for the differences between device and observed systolic and diastolic blood pressure values fulfilled both criterion 1 and criterion 2 of the standard (criterion 1: -2.62±6.89 and -0.90±6.63 mmHg, respectively; criterion 2: -2.65±5.36 and -0.91±6.12 mmHg, respectively). The Bland-Altman plots did not show any systematic variation in the error.

CONCLUSION:

The UA-1200BLE device had a high level of accuracy and fulfilled all ISO 81060-2:2013 standard requirements in adults with a wide range of arm circumference measurements. It is therefore suitable as a wireless blood pressure monitoring tool in patients with hypertension.