Jpn J Ophthalmol. 2018 Jul;62(4):443-450. doi: 10.1007/s10384-018-0594-5. Epub 2018 Apr 26.

Evaluation of corneal biomechanics in patients with keratectasia following LASIK using dynamic Scheimpflug analyzer.

Author information

1
Department of Ophthalmology, Osaka University Graduate School of Medicine, Room E7, 2-2 Yamadaoka, Suita, Osaka, 565-0871, Japan.
2
Department of Ophthalmology and Visual Sciences, Graduate School of Biomedical Sciences, Nagasaki University, Nagasaki, Japan.
3
Department of Ophthalmology, Osaka University Graduate School of Medicine, Room E7, 2-2 Yamadaoka, Suita, Osaka, 565-0871, Japan. nmaeda@ophthal.med.osaka-u.ac.jp.

Abstract

PURPOSE:

To investigate the corneal biomechanics in eyes with keratectasia following LASIK using a dynamic Scheimpflug analyzer.

DESIGN:

Case-Control study.

METHOD:

The subjects in the study included 12 eyes with keratectasia after LASIK (KE), 24 eyes with keratoconus (KC), 17 eyes without keratectasia after LASIK (LASIK), and 34 eyes with normal corneas (Normal). Corneal biomechanics of the four groups were evaluated using a dynamic Scheimpflug analyzer.

RESULTS:

Compared with Normal (7.06 ± 0.54), the radius at the highest concavity (radius, mm) of LASIK (5.96 ± 0.76), KE (4.93 ± 0.61) and KC (5.39 ± 1.02) were significantly small. The Deflection Amplitude (HCDLA, mm) of Normal (0.94 ± 0.07) was significantly lower than those of KE (1.11 ± 0.10) and KC (1.06 ± 0.16), and was not significantly different from that of LASIK (0.98 ± 0.07). There were significant differences between LASIK and KE in radius and HCDLA (P < 0.05), whereas KE and KC had no differences in these parameters.

CONCLUSIONS:

Corneal biomechanical features evaluated using the dynamic Scheimpflug analyzer suggest that biomechanical properties in eyes with keratectasia, keratoconus, and LASIK are different from those of normal eyes. Although the biomechanics in eyes with keratectasia differs from that in eyes with LASIK, it is similar to that in eyes with keratoconus.

KEYWORDS:

Biomechanics; Dynamic Scheimpflug analyzer; Keratectasia; Keratoconus; LASIK