Regeneration of Arrayed Gold Microelectrodes Equipped for a Real-Time Cell Analyzer

J Vis Exp. 2018 Mar 12:(133):56250. doi: 10.3791/56250.

Abstract

The label-free cell-based assay is advantageous for biochemical study because of it does not require the use of experimental animals. Due to its ability to provide more dynamic information about cells under physiological conditions than classical biochemical assays, this label-free real-time cell assay based on the electric impedance principle is attracting more attention during the past decade. However, its practical utilization may be limited due to the relatively expensive cost of measurement, in which costly consumable disposable gold microchips are used for the cell analyzer. In this protocol, we have developed a general strategy to regenerate arrayed gold microelectrodes equipped for a commercial label-free cell analyzer. The regeneration process includes trypsin digestion, rinsing with ethanol and water, and a spinning step. The proposed method has been tested and shown to be effective for the regeneration and repeated usage of commercial electronic plates at least three times, which will help researchers save on the high running cost of real-time cell assays.

Publication types

  • Research Support, Non-U.S. Gov't
  • Video-Audio Media

MeSH terms

  • A549 Cells
  • Animals
  • Antineoplastic Agents / pharmacology
  • Computer Systems
  • Cytological Techniques / instrumentation*
  • Cytological Techniques / methods
  • Drug Screening Assays, Antitumor / instrumentation
  • Drug Screening Assays, Antitumor / methods
  • Electric Impedance
  • Gold / chemistry*
  • Humans
  • Microelectrodes*

Substances

  • Antineoplastic Agents
  • Gold