Clin Biochem. 2018 May;55:93-95. doi: 10.1016/j.clinbiochem.2018.03.011. Epub 2018 Mar 16.

Blood gas sample spiking with total parenteral nutrition, lipid emulsion, and concentrated dextrose solutions as a model for predicting sample contamination based on glucose result.

Author information

1
Department of Laboratory Medicine and Pathology, Mayo Clinic, 200 First Street SW, Rochester, MN 55905, United States.
2
Department of Laboratory Medicine and Pathology, Mayo Clinic, 200 First Street SW, Rochester, MN 55905, United States. Electronic address: Karon.bradley@mayo.edu.

Abstract

OBJECTIVE:

Evaluate the effects of blood gas sample contamination with total parenteral nutrition (TPN)/lipid emulsion and dextrose 50% (D50) solutions on blood gas and electrolyte measurement; and determine whether glucose concentration can predict blood gas sample contamination with TPN/lipid emulsion or D50.

DESIGN AND METHODS:

Residual lithium heparin arterial blood gas samples were spiked with TPN/lipid emulsion (0 to 15%) and D50 solutions (0 to 2.5%). Blood gas (pH, pCO2, pO2), electrolytes (Na+, K+ ionized calcium) and hemoglobin were measured with a Radiometer ABL90. Glucose concentration was measured in separated plasma by Roche Cobas c501. Chart review of neonatal blood gas results with glucose >300 mg/dL (>16.65 mmol/L) over a seven month period was performed to determine whether repeat (within 4 h) blood gas results suggested pre-analytical errors in blood gas results. Results were used to determine whether a glucose threshold could predict contamination resulting in blood gas and electrolyte results with greater than laboratory-defined allowable error.

RESULTS:

Samples spiked with 5% or more TPN/lipid emulsion solution or 1% D50 showed glucose concentration >500 mg/dL (>27.75 mmol/L) and produced blood gas (pH, pO2, pCO2) results with greater than laboratory-defined allowable error. TPN/lipid emulsion, but not D50, produced greater than allowable error in electrolyte (Na+,K+,Ca++,Hb) results at these concentrations. Based on chart review of 144 neonatal blood gas results with glucose >250 mg/dL received over seven months, four of ten neonatal intensive care unit (NICU) patients with glucose results >500 mg/dL and repeat blood gas results within 4 h had results highly suggestive of pre-analytical error. Only 3 of 36 NICU patients with glucose results 300-500 mg/dL and repeat blood gas results within 4 h had clear pre-analytical errors in blood gas results.

CONCLUSION:

Glucose concentration can be used as an indicator of significant blood sample contamination with either TPN/lipid emulsion or D50 solution. NICU blood gas samples with glucose ≥300 mg/dL should be considered potentially contaminated, and samples with glucose >500 mg/dL have a risk for contamination.

KEYWORDS:

Blood gas analysis; Dextrose; Interferences; Lipid emulsion; Pre analytical errors; Total parenteral nutrition

[Indexed for MEDLINE]