Format

Send to

Choose Destination
Arch Dis Child Fetal Neonatal Ed. 2019 Mar;104(2):F187-F191. doi: 10.1136/archdischild-2017-313982. Epub 2018 Mar 17.

Detection of exhaled carbon dioxide following intubation during resuscitation at delivery.

Author information

1
MRC & Asthma UK Centre in Allergic Mechanisms in Asthma, King's College London, London, UK.
2
Department of Women and Children's Health, School of Life Course Sciences, Faculty of Life Sciences and Medicine, King's College London, London, UK.
3
Neonatal Unit, Southern General and Yorkhill Hospitals, Scotland, UK.
4
Evelina Children's Hospital Neonatal Unit, Guy's and St Thomas' Hospital NHS Foundation Trust, London, UK.
5
National Institute for Health Research (NIHR) Biomedical Research Centre, Guy's and St Thomas' NHS Foundation Trust, King's College London, London, UK.

Abstract

OBJECTIVES:

End tidal carbon dioxide (ETCO2) monitoring can facilitate identification of successful intubation. The aims of this study were to determine the time to detect ETCO2 following intubation during resuscitation of infants born prematurely and whether it differed according to maturity at birth or the Apgar scores (as a measure of the infant's condition after birth).

DESIGN:

Analysis of recordings of respiratory function monitoring.

SETTING:

Two tertiary perinatal centres.

PATIENTS:

Sixty-four infants, with median gestational age of 27 (range 23-34)weeks.

INTERVENTIONS:

Respiratory function monitoring during resuscitation in the delivery suite.

MAIN OUTCOME MEASURES:

The time following intubation for ETCO2 levels to be initially detected and to reach 4 mm Hg and 15 mm Hg.

RESULTS:

The median time for initial detection of ETCO2 following intubation was 3.7 (range 0-44) s, which was significantly shorter than the median time for ETCO2 to reach 4 mm Hg (5.3 (range 0-727) s) and to reach 15 mm Hg (8.1 (range 0-827) s) (both P<0.001). There were significant correlations between the time for ETCO2 to reach 4 mm Hg (r=-0.44, P>0.001) and 15 mm Hg (r=-0.48, P<0.001) and gestational age but not with the Apgar scores.

CONCLUSIONS:

The time for ETCO2 to be detected following intubation in the delivery suite is variable emphasising the importance of using clinical indicators to assess correct endotracheal tube position in addition to ETCO2 monitoring. Capnography is likely to detect ETCO2 faster than colorimetric devices.

KEYWORDS:

end tidal carbon dioxide; intubation; resuscitation