Clinical performance of a novel textile interface for neonatal chest electrical impedance tomography

Physiol Meas. 2018 Apr 26;39(4):044004. doi: 10.1088/1361-6579/aab513.

Abstract

Objective: Critically ill neonates and infants might particularly benefit from continuous chest electrical impedance tomography (EIT) monitoring at the bedside. In this study a textile 32-electrode interface for neonatal EIT examination has been developed and tested to validate its clinical performance. The objectives were to assess ease of use in a clinical setting, stability of contact impedance at the electrode-skin interface and possible adverse effects.

Approach: Thirty preterm infants (gestational age: 30.3 ± 3.9 week (mean ± SD), postnatal age: 13.8 ± 28.2 d, body weight at inclusion: 1727 ± 869 g) were included in this multicentre study. The electrode-skin contact impedances were measured continuously for up to 3 d and analysed during the initial 20-min phase after fastening the belt and during a 10 h measurement interval without any clinical interventions. The skin condition was assessed by attending clinicians.

Main results: Our findings imply that the textile electrode interface is suitable for long-term neonatal chest EIT imaging. It does not cause any distress for the preterm infants or discomfort. Stable contact impedance of about 300 Ohm was observed immediately after fastening the electrode belt and during the subsequent 20 min period. A slight increase in contact impedance was observed over time. Tidal variation of contact impedance was less than 5 Ohm.

Significance: The availability of a textile 32-electrode belt for neonatal EIT imaging with simple, fast, accurate and reproducible placement on the chest strengthens the potential of EIT to be used for regional lung monitoring in critically ill neonates and infants.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts
  • Electric Impedance
  • Electrodes
  • Humans
  • Infant, Newborn
  • Skin
  • Surface Properties
  • Textiles*
  • Thorax / diagnostic imaging*
  • Tomography / instrumentation*