Diabetes Technol Ther. 2018 Mar;20(3):180-188. doi: 10.1089/dia.2017.0386. Epub 2018 Feb 22.

Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes.

Author information

1
1 Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, United Kingdom .
2
2 School of Medicine, Newcastle University Medicine Malaysia , Johor, Malaysia .
3
3 Gender Medicine Unit, Division of Endocrinology and Metabolism, Department of Internal Medicine III, Medical University of Vienna, Vienna, Austria; and Gender Medicine Institute, Gars am Kamp, Austria .

Abstract

BACKGROUND:

Accuracy of the FreeStyle Libre™ Flash Glucose Monitoring System has not been evaluated in pregnant women with diabetes. The aim of this study was to determine accuracy (compared to self-monitoring of blood glucose [SMBG]), clinical safety, and acceptability of the FreeStyle Libre System when used at home by this population.

MATERIALS AND METHODS:

Seventy-four participants, with type 1 (T1D, n = 24), type 2 (T2D, n = 11), or gestational (n = 39) diabetes, were enrolled across 13 sites (9 in United Kingdom, 4 in Austria). Average gestation was 26.6 ± 6.8 weeks (mean ± standard deviation), age was 30.5 ± 5.1 years, diabetes duration was 13.1 ± 7.3 years for T1D and 3.2 ± 2.5 years for T2D, and 49/74 (66.2%) used insulin to manage their diabetes. Sensors were worn for up to 14 days. Sensor glucose values (masked) were compared with capillary SMBG values (made at least 4 times/day).

RESULTS:

Clinical accuracy of sensor results versus SMBG results was demonstrated, with 88.1% and 99.8% of results within Zone A and Zones A and B of the Consensus Error Grid, respectively. Overall mean absolute relative difference was 11.8%. Sensor accuracy was unaffected by the type of diabetes, the stage of pregnancy, whether insulin was used, age or body mass index. User questionnaires indicated high levels of satisfaction with sensor wear, system use, and comparison to SMBG. There were no unanticipated device-related adverse events.

CONCLUSIONS:

Good agreement was demonstrated between the FreeStyle Libre System and SMBG. Accuracy of the system was unaffected by patient characteristics, indicating that the system is safe and accurate to use by pregnant women with diabetes.

KEYWORDS:

Accuracy.; Diabetes; Flash glucose monitoring; Pregnancy; Sensor