Vet Clin Pathol. 2018 Mar;47(1):9-21. doi: 10.1111/vcp.12583. Epub 2018 Feb 11.

ASVCP guidelines: Allowable total error hematology.

Author information

1
Texas A&M University, College Station, TX, USA.
2
Urika, LLC, Mukilteo, WA, USA.
3
University of Georgia, Athens, GA, USA.
4
University of Tennessee, Knoxville, TN, USA.
5
Colorado State University, Fort Collins, CO, USA.

Abstract

The purpose of this document is to provide total allowable error (TEa ) recommendations for commonly analyzed hematology measurands for veterinary personnel. These guidelines define relevant terminology and highlight considerations specific to hematology measurands. They also provide reasons and guidelines for using TEa in instrument performance evaluation, including recommendations for when the total observed error exceeds the recommended TEa . Biological variation-based quality specifications are briefly discussed. The appendix describes the derivation of the hematology TEa recommendations and provides resources for external quality assurance/proficiency testing programs and a worksheet for implementation of the guidelines.

KEYWORDS:

Bias; Biological variation; Imprecision; Instrument performance; Quality assurance; Quality control; Total observed error