Food Nutr Bull. 2018 Mar;39(1):28-38. doi: 10.1177/0379572117742502. Epub 2017 Dec 19.

First Field Test of an Innovative, Wider Tape to Measure Mid-Upper Arm Circumference in Young Laotian Children.

Author information

1
1 Program in International and Community Nutrition, Department of Nutrition, University of California, Davis, Davis, CA, USA.
2
2 National Institute of Public Health, Vientiane, Lao People's Democratic Republic.
3
3 Intellectual Ventures Global Good Fund, Bellevue, WA, USA.

Abstract

BACKGROUND:

Measurements of mid-upper arm circumference (MUAC) may result in measurement error due to incorrect placement along the arm or tight pulling of tape. To reduce the risk of these measurement errors, a new wider tape was developed.

OBJECTIVE:

To compare the measurement agreement and precision and the ease of use of the standard and wide MUAC tapes.

METHODS:

Mid-upper arm circumference was measured in 814 children aged 9 to 32 months with both tapes. The midpoint of the upper arm was measured with the standard tape and estimated with the wide tape. Standardization sessions were implemented to assess intra- and interobserver precision.

RESULTS:

Mid-upper arm circumference with the wide MUAC tape was significantly larger than the standard tape (mean [standard deviation]: 14.3 [1.0] cm vs 13.9 [1.0] cm; P < .001), resulting in a consistent bias of +0.41 cm. Forty-six (5.7%) children were identified with low MUAC <12.5 cm by standard tape compared with 10 (1.2%) by the wide tape ( P <.001). Because a new tape could be reproduced by correcting for this bias, we corrected measured results by subtracting 0.41 cm and mean MUAC by tape type was no longer significantly different. Intra- and interobserver technical error of measurement suggested a better precision with the wide MUAC tape.

CONCLUSIONS:

Despite simplifying the measurement by approximating the midpoint of the upper arm, the wide MUAC tape tended to have better precision than the standard MUAC tape. However, there was a consistent measurement bias of +0.41 cm in mean MUAC. This first field test yielded promising results and led to further product adjustments.

KEYWORDS:

Lao PDR; MUAC; mid-upper arm circumference; young children

PMID:
29258337
DOI:
10.1177/0379572117742502
[Indexed for MEDLINE]