Nature. 2017 Nov 8;551(7679):232-236. doi: 10.1038/nature24636.

Fully integrated silicon probes for high-density recording of neural activity.

Author information

1
HHMI Janelia Research Campus, 19700 Helix Drive, Ashburn, Virginia 20147, USA.
2
UCL Institute of Neurology, University College London, London WC1N 3BG, UK.
3
Department of Neuroscience, Physiology and Pharmacology, University College London, London WC1E 6DE, UK.
4
UCL Institute of Ophthalmology, University College London, London EC1V 9EL, UK.
5
Allen Institute for Brain Science, 615 Westlake Avenue North, Seattle, Washington 98109, USA.
6
Department of Cell and Developmental Biology, University College London, London WC1E 6BT, UK.
7
Sainsbury Wellcome Centre, University College London, London W1T 4JG, UK.
8
Department of Neurology, University of British Columbia, Vancouver, British Columbia V6T 2B5, Canada.
9
imec, Kapeldreef 75, 3001 Heverlee, Leuven, Belgium.
10
Neuro-Electronics Research Flanders, Kapeldreef 75, 3001 Leuven, Belgium.
11
KU Leuven, Department of Biology, Naamsestraat 59, 3000 Leuven, Belgium.
12
White Matter LLC, 999 3rd Avenue 700, 98104 Seattle, USA.
13
VIB, 3001 Leuven, Belgium.
14
Wolfson Institute for Biomedical Research, University College London, Gower Street, London WC1E 6BT, UK.
15
Centre for Systems Neuroscience, University of Leicester, Leicester LE1 7QR, UK.

Abstract

Sensory, motor and cognitive operations involve the coordinated action of large neuronal populations across multiple brain regions in both superficial and deep structures. Existing extracellular probes record neural activity with excellent spatial and temporal (sub-millisecond) resolution, but from only a few dozen neurons per shank. Optical Ca2+ imaging offers more coverage but lacks the temporal resolution needed to distinguish individual spikes reliably and does not measure local field potentials. Until now, no technology compatible with use in unrestrained animals has combined high spatiotemporal resolution with large volume coverage. Here we design, fabricate and test a new silicon probe known as Neuropixels to meet this need. Each probe has 384 recording channels that can programmably address 960 complementary metal-oxide-semiconductor (CMOS) processing-compatible low-impedance TiN sites that tile a single 10-mm long, 70 × 20-μm cross-section shank. The 6 × 9-mm probe base is fabricated with the shank on a single chip. Voltage signals are filtered, amplified, multiplexed and digitized on the base, allowing the direct transmission of noise-free digital data from the probe. The combination of dense recording sites and high channel count yielded well-isolated spiking activity from hundreds of neurons per probe implanted in mice and rats. Using two probes, more than 700 well-isolated single neurons were recorded simultaneously from five brain structures in an awake mouse. The fully integrated functionality and small size of Neuropixels probes allowed large populations of neurons from several brain structures to be recorded in freely moving animals. This combination of high-performance electrode technology and scalable chip fabrication methods opens a path towards recording of brain-wide neural activity during behaviour.