A reusable OSL-film for 2D radiotherapy dosimetry

Phys Med Biol. 2017 Oct 19;62(21):8441-8454. doi: 10.1088/1361-6560/aa8de6.

Abstract

Optical stimulated luminescence (OSL) combines reusability, sub-mm resolution, and a linear dose response in a single radiation detection technology. Such a combination is currently lacking in radiotherapy dosimetry. But OSL-films have a strong energy dependent response to keV photons due to a relative high effective atomic number (Z eff). The current work studied the applicability of a 2D OSL-film with a reduced Z eff as (IMRT/VMAT) dosimeter. Based on their commercial OSL-film experience, Agfa Healthcare N.V. produced a new experimental OSL-film for RT dosimetry. This film had a lower effective atomic number compared to the films used in radiology. Typical 2D dosimeter requirements such as uniformity, dose response, signal stability with time, and angular dependence were evaluated. Additionally, the impact of a possible residual energy dependence was assessed for the infield as well as the out-of-field region of both static beams and standard intensity modulated patterns (chair and pyramid). The OSL-film's reusable nature allowed for a film specific absolute and linear calibration including a flood-field uniformity correction. The OSL-film was scanned with a CR-15X engine based reader using a strict timing (i.e. 4 min after 'beam on' or as soon as possible) to account for spontaneous recombination. The OSL-film had good basic response properties: non-uniformities ⩽2.6%, a linear dose response (0-32 Gy), a linear signal decay (0.5% min-1) over the 20 min measured, and limited angular dependence ⩽2.6%. Due to variations of the energy spectrum, larger dose differences were noted outside the central region of the homogenous phantom and outside both static and IMRT fields. However, the OSL-film's measured dose differences of the IMRT patterns were lower than those of Gafchromic EBT measurements ([-1.6%, 2.1%] versus [-2.9%, 3.6%]). The current OSL-film could be used as a reusable high resolution dosimeter with read-out immediately after irradiation. Inside the (IMRT) treatment fields residual energy dependent effects were not observed. Novelty and significance: Implementing a reusable optical stimulated luminescence (OSL) film for radiotherapy dosimetry would enable user-friendly, sub(mm) resolution 2D dosimetry with instantaneous read-out. Radiology OSL-films have a strong energy dependent response which hampers accurate dosimetry. The current work reports measurements with a first 2D OSL-film tailored to the radiotherapy needs: including an improved water equivalent composition. The dosimeter adds the ability for sub-mm resolution repeated measurements to the portfolio of radiotherapy dosimetry.

MeSH terms

  • Calibration
  • Equipment Design
  • Film Dosimetry / instrumentation*
  • Film Dosimetry / methods
  • Humans
  • Luminescence*
  • Materials Testing*
  • Phantoms, Imaging*
  • Photons
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Water / chemistry

Substances

  • Water