Impact of targeted education on managing warning and error signals by children and adolescents with type 1 diabetes using the Accu-Chek Combo Insulin Pump System

Ann Pediatr Endocrinol Metab. 2017 Sep;22(3):170-175. doi: 10.6065/apem.2017.22.3.170. Epub 2017 Sep 28.

Abstract

Purpose: Insulin pumps are widely used in diabetes. They are equipped with safety alarms to alert users. Pump manuals contain alarm codes and how to troubleshoot them. However, these manuals are lengthy and difficult to use, particularly in emergencies. We aim to assess the impact of targeted education on warnings and errors in improving competency to troubleshoot the alarms.

Methods: Twenty-one patients, with a median age of 13, were recruited over a 5-month period. Each patient had 2 study visits. The frequencies and types of alarms were recorded, and patients were given a summary sheet that outlined common alarms encountered and troubleshooting tips. In visit 2, the frequencies and types of alarms were compared to those of visit 1. The patients were asked to fill a questionnaire and to rate the education session given in visit 1, their level of competency in decrypting alarm codes, and their promptness in responding to alarms.

Results: Low cartridge (W1), low battery (W2), and bolus cancelled (W8) were the commonest warnings. The most noted errors were occlusion (E4), power interruption (E8), empty battery (E2), set not primed (E11), and cartridge empty (E1). The numbers of warning and error signals markedly decreased after targeted education (P<0.05). The ability in decrypting warning signals significantly improved (P=0.02), and the frequency of response to pump alarms significantly increased (P=0.001).

Conclusions: Certain warnings and errors are more common than others in insulin pumps. Targeted education is useful in improving competency and response of patients in managing pump alarms.

Keywords: Clinical alarms; Insulin infusion systems; Medical errors; Type 1 diabetes mellitus; Warnings.