Design and development of the BD Odon DeviceTM : a human factors evaluation process.

Author information

1
Department of Obstetrics and Gynaecology, North Bristol NHS Trust, Southmead Hospital, Bristol, UK.
2
School of Clinical Sciences, University of Bristol, Bristol, UK.
3
Preanalytical Systems, BD, Franklin Lakes, NJ, USA.
4
Inventor of BD Odon Device, Banfield, Buenos Aires, Argentina.
5
School of Social and Community Medicine, University of Bristol, Bristol, UK.
6
R&D Department, BD, Singapore, Singapore.
7
Global Health, BD, Franklin Lakes, NJ, USA.
8
Corporate Affairs, BD, Franklin Lakes, NJ, USA.

Abstract

OBJECTIVE:

To (1) determine how intended users interact with and use the BD Odon Device in simulation, (2) use these findings to alter progressively the design of the BD Odon Device and (3) validate that these changes have improved the ability of practitioners to use the BD Odon Device.

DESIGN:

Human factors evaluation study.

SETTING:

Simulation suite designed to mimic delivery room.

POPULATION OR SAMPLE:

Three hundred and ninety simulated operative births, performed by 100 practising clinicians.

METHODS:

Simulated operative vaginal births performed using the BD Odon Device and the device Instructions for use were subjected to three formative human factors evaluations and one human factors validation test. Following each evaluation, findings were reviewed and the design of the BD Odon Device and Instructions for use were modified.

MAIN OUTCOME MEASURES:

Successful performance of an operative vaginal birth using the BD Odon Device in accordance with provided training and Instructions for use.

RESULTS:

Using version two of the BD Odon Device, and following exposure to face-to-face training and written instructions, 25% of accouchers were able successfully to perform a simulated operative vaginal birth. In the final evaluation, following device design and training material alterations, all accouchers were able successfully to perform a simulated operative vaginal birth using version four of the BD Odon Device.

CONCLUSIONS:

Human factors evaluations have enabled a multi-professional device and training materials design team to alter the design of the BD Odon Device and the Instructions for use in an evidence-based fashion. This process has resulted in a device which has a predictable and likely safe pattern of use.

TWEETABLE ABSTRACT:

Human Factors evaluations help make the BD Odon Device safe and usable for clinical practice.

KEYWORDS:

BD Odon Device; human factors; improvements; use