Effects of Flexible Dry Electrode Design on Electrodermal Activity Stimulus Response Detection

IEEE Trans Biomed Eng. 2017 Dec;64(12):2979-2987. doi: 10.1109/TBME.2017.2754220. Epub 2017 Sep 18.

Abstract

Objective: The focus of this research is to evaluate the effects of design parameters including surface area, distance between and geometry of dry flexible electrodes on electrodermal activity (EDA) stimulus response detection.

Methods: EDA is a result of the autonomic nervous system being stimulated, which causes sweat and changes the electrical characteristics of the skin. Standard silver/silver chloride (Ag/AgCl) EDA electrodes are rigid and lack conformability in contact with skin. In this study, flexible dry Ag/AgCl EDA electrodes were fabricated on a compliant substrate, used to monitor EDA stimulus responses and compared to results simultaneously collected by rigid dry Ag/AgCl electrodes.

Results: A repeatable fabrication process for flexible Ag/AgCl electrodes has been established. Surface area, distance between and geometry of electrodes are shown to affect the detectability of the EDA response and the minimum number of sweat glands to be covered by the electrodes has been estimated at 140, or more, in order to maintain functionality. The optimal flexible EDA electrode is a serpentine design with a 0.15 cm2 surface area and a 0.20 cm distance with an average Pearson correlation coefficient of .

Conclusion: Fabrication of flexible electrodes is described and an understanding of the effects of electrode designs on the EDA stimulus response detection has been established and is potentially related to the coverage of sweat glands.

Significance: This work presents a novel systematic approach to understand the effects of electrode designs on monitoring EDA which is of importance for the design of wearable EDA monitoring devices.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Clothing
  • Electrodes
  • Equipment Design
  • Galvanic Skin Response / physiology*
  • Humans
  • Monitoring, Ambulatory / instrumentation*
  • Monitoring, Ambulatory / methods
  • Signal Processing, Computer-Assisted / instrumentation*