Invest Ophthalmol Vis Sci. 2017 Jul 1;58(9):3608-3615. doi: 10.1167/iovs.17-21904.

Reliability and Repeatability of Cone Density Measurements in Patients With Stargardt Disease and RPGR-Associated Retinopathy.

Author information

1
UCL Institute of Ophthalmology, University College London, London, United Kingdom 2Moorfields Eye Hospital, London, United Kingdom.
2
UCL Institute of Ophthalmology, University College London, London, United Kingdom 2Moorfields Eye Hospital, London, United Kingdom 3Department of Ophthalmology, Medical University Graz and Johannes Kepler University, Linz, Austria 4Department of Ophthalmology, University of Basel, Basel, Switzerland.
3
Division of Biostatistics, Institute for Health and Society, Medical College of Wisconsin, Milwaukee, Wisconsin, United States.
4
Byers Eye Institute, Stanford University, Palo Alto, California, United States.
5
Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, Wisconsin, United States 8Department of Cell Biology, Neurobiology and Anatomy, Medical College of Wisconsin, Milwaukee, Wisconsin, United States 9Department of Biophysics, Medical College of Wisconsin, Milwaukee, Wisconsin, United States.

Abstract

Purpose:

To assess reliability and repeatability of cone density measurements by using confocal and (nonconfocal) split-detector adaptive optics scanning light ophthalmoscopy (AOSLO) imaging. It will be determined whether cone density values are significantly different between modalities in Stargardt disease (STGD) and retinitis pigmentosa GTPase regulator (RPGR)-associated retinopathy.

Methods:

Twelve patients with STGD (aged 9-52 years) and eight with RPGR-associated retinopathy (aged 11-31 years) were imaged using both confocal and split-detector AOSLO simultaneously. Four graders manually identified cone locations in each image that were used to calculate local densities. Each imaging modality was evaluated independently. The data set consisted of 1584 assessments of 99 STGD images (each image in two modalities and four graders who graded each image twice) and 928 RPGR assessments of 58 images (each image in two modalities and four graders who graded each image twice).

Results:

For STGD assessments the reliability for confocal and split-detector AOSLO was 67.9% and 95.9%, respectively, and the repeatability was 71.2% and 97.3%, respectively. The differences in the measured cone density values between modalities were statistically significant for one grader. For RPGR assessments the reliability for confocal and split-detector AOSLO was 22.1% and 88.5%, respectively, and repeatability was 63.2% and 94.5%, respectively. The differences in cone density between modalities were statistically significant for all graders.

Conclusions:

Split-detector AOSLO greatly improved the reliability and repeatability of cone density measurements in both disorders and will be valuable for natural history studies and clinical trials using AOSLO. However, it appears that these indices may be disease dependent, implying the need for similar investigations in other conditions.

PMID:
28738413
PMCID:
PMC5525557
DOI:
10.1167/iovs.17-21904
[Indexed for MEDLINE]