Current microfluidic assays, which aim at quantifying mechanical properties of sickle cell red blood cells (SS-RBCs), suffer from a number of drawbacks in functionalization and flow control. Specifically, physical adsorption functionalization techniques produce inconsistent functional surfaces, and common volumetric flow pumps cannot be used to adjust the flow inside microchannels with minimal delay. We have designed an experimental setup that alleviates these complications by implementing aspiration for microchannel assembly that enables the use of most functionalization techniques and a pressure controller that allows instant and precise changes in the microchannel flow. Utilizing this setup, we have quantified SS-RBC adhesion to the integrin αvβ3, a specific adhesion protein expressed on the endothelium, as well as measured the shear modulus and viscosity of the SS-RBC plasma membrane.

KEYWORDS:

Erythrocytes; ICAM-4; Neo-Hookean; Surface functionalization; Viscoelasticity; αvβ3

PMID:
28267661
DOI:
10.1016/j.jmbbm.2017.02.031
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Elsevier Science
Loading ...
Support Center