Optical See-Through Cancer Vision Goggles Enable Direct Patient Visualization and Real-Time Fluorescence-Guided Oncologic Surgery

Ann Surg Oncol. 2017 Jul;24(7):1897-1903. doi: 10.1245/s10434-017-5804-8. Epub 2017 Feb 17.

Abstract

Background: The inability to visualize the patient and surgical site directly, limits the use of current near infrared fluorescence-guided surgery systems for real-time sentinel lymph node biopsy and tumor margin assessment.

Methods: We evaluated an optical see-through goggle augmented imaging and navigation system (GAINS) for near-infrared, fluorescence-guided surgery. Tumor-bearing mice injected with a near infrared cancer-targeting agent underwent fluorescence-guided, tumor resection. Female Yorkshire pigs received hind leg intradermal indocyanine green injection and underwent fluorescence-guided, popliteal lymph node resection. Four breast cancer patients received 99mTc-sulfur colloid and indocyanine green retroareolarly before undergoing sentinel lymph node biopsy using radioactive tracking and fluorescence imaging. Three other breast cancer patients received indocyanine green retroareolarly before undergoing standard-of-care partial mastectomy, followed by fluorescence imaging of resected tumor and tumor cavity for margin assessment.

Results: Using near-infrared fluorescence from the dyes, the optical see-through GAINS accurately identified all mouse tumors, pig lymphatics, and four pig popliteal lymph nodes with high signal-to-background ratio. In 4 human breast cancer patients, 11 sentinel lymph nodes were identified with a detection sensitivity of 86.67 ± 0.27% for radioactive tracking and 100% for GAINS. Tumor margin status was accurately predicted by GAINS in all three patients, including clear margins in patients 1 and 2 and positive margins in patient 3 as confirmed by paraffin-embedded section histopathology.

Conclusions: The optical see-through GAINS prototype enhances near infrared fluorescence-guided surgery for sentinel lymph node biopsy and tumor margin assessment in breast cancer patients without disrupting the surgical workflow in the operating room.

MeSH terms

  • Adult
  • Aged
  • Animals
  • Breast Neoplasms / pathology
  • Breast Neoplasms / surgery*
  • Eye Protective Devices*
  • Female
  • Fluorescence*
  • Humans
  • Indocyanine Green
  • Lymph Node Excision
  • Lymph Nodes / pathology
  • Lymph Nodes / surgery*
  • Mice
  • Mice, Inbred BALB C
  • Mice, Nude
  • Middle Aged
  • Prognosis
  • Sentinel Lymph Node Biopsy
  • Surgery, Computer-Assisted / methods*
  • Surgical Oncology*
  • Swine

Substances

  • Indocyanine Green