Catheter Cardiovasc Interv. 2017 May;89(6):1045-1050. doi: 10.1002/ccd.26860. Epub 2016 Nov 10.

Single-center experience in percutaneous closure of arterial duct with Amplatzer duct Occluder II additional sizes.

Author information

1
Department of Paediatric Cardiology, A.O.R.N. "Ospedali dei Colli", 2nd University of Naples, Naples, Italy.

Abstract

OBJECTIVES:

This study aimed to report a large, single-center experience of percutaneous arterial duct (AD) closure using Amplatzer Duct Occluder II Additional Sizes device (ADO II-AS)(St. Jude Medical Corp, St. Paul, MN, USA).

BACKGROUND:

Transcatheter closure of AD remains challenging in low body weight patients and those who have a persisting shunt following a previous attempt at interventional closure. Recent technical advances in device design may address these issues.

METHODS:

From May 2011 to April 2016, 109 patients underwent attempted percutaneous closure of AD with ADO II-AS at our Institution. Mean age and weight were 4.8 ± 8.1 years (range 0-48) and 21.4 ± 20.6 kg (range 3-93), respectively. Fifteen patients (13.8%) were ≤6 kg (age 3.5 ± 2.0 months; weight 4.7 ± 1.1 kg). Arterial duct morphology was type A in 62 (57%), type B in 1 (1%), type C in 32 (29%), type D in 7 (6%) and type E in 6 patients (6%), respectively. Arterial approach was used to negotiate and deploy the occluding device in 103 patients (94.5%).

RESULTS:

AD diameter was 2.2 ± 0.6 (range 1.5-4.5) resulting in QP/QS of 1.9 ± 0.7 (range 1-3.3). Mean pulmonary artery pressure and PA/aortic pressure ratio were 19.3 ± 5.0 mm Hg (range 12-38) and 0.34 ± 0.14 (range 0.14-0.95), respectively. Successful device deployment was achieved in 107 patients (98.2%). Neither procedural morbidity nor mortality was recorded. Immediate, 24h and mid-term (30 ± 17 months) complete occlusion was recorded in 71%, 98.1%, and 100% of patients, respectively.

CONCLUSION:

In our experience, trans-catheter closure of AD of different sizes and morphologies using ADO II-AS is highly feasible, safe and effective also in challenging anatomic/clinical settings. © 2016 Wiley Periodicals, Inc.

KEYWORDS:

ASD/PDA/PFO; closure; congenital heart disease; pediatric intervention; pediatrics