J Am Heart Assoc. 2016 Jul 21;5(7). pii: e003428. doi: 10.1161/JAHA.116.003428.

Diagnostic Performance of a Smartphone-Based Photoplethysmographic Application for Atrial Fibrillation Screening in a Primary Care Setting.

Author information

1
Cardiology Division, Department of Medicine, Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong, China.
2
Cardiio Inc., Cambridge, MA.
3
Department of Family Medicine and Primary Healthcare, Hong Kong East Cluster, Hong Kong, China.
4
Cardiio Inc., Cambridge, MA cwdsiu@hku.hk mingzher@cardiio.com.
5
Cardiology Division, Department of Medicine, Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong, China cwdsiu@hku.hk mingzher@cardiio.com.

Abstract

BACKGROUND:

Diagnosing atrial fibrillation (AF) before ischemic stroke occurs is a priority for stroke prevention in AF. Smartphone camera-based photoplethysmographic (PPG) pulse waveform measurement discriminates between different heart rhythms, but its ability to diagnose AF in real-world situations has not been adequately investigated. We sought to assess the diagnostic performance of a standalone smartphone PPG application, Cardiio Rhythm, for AF screening in primary care setting.

METHODS AND RESULTS:

Patients with hypertension, with diabetes mellitus, and/or aged ≥65 years were recruited. A single-lead ECG was recorded by using the AliveCor heart monitor with tracings reviewed subsequently by 2 cardiologists to provide the reference standard. PPG measurements were performed by using the Cardiio Rhythm smartphone application. AF was diagnosed in 28 (2.76%) of 1013 participants. The diagnostic sensitivity of the Cardiio Rhythm for AF detection was 92.9% (95% CI] 77-99%) and was higher than that of the AliveCor automated algorithm (71.4% [95% CI 51-87%]). The specificities of Cardiio Rhythm and the AliveCor automated algorithm were comparable (97.7% [95% CI: 97-99%] versus 99.4% [95% CI 99-100%]). The positive predictive value of the Cardiio Rhythm was lower than that of the AliveCor automated algorithm (53.1% [95% CI 38-67%] versus 76.9% [95% CI 56-91%]); both had a very high negative predictive value (99.8% [95% CI 99-100%] versus 99.2% [95% CI 98-100%]).

CONCLUSIONS:

The Cardiio Rhythm smartphone PPG application provides an accurate and reliable means to detect AF in patients at risk of developing AF and has the potential to enable population-based screening for AF.

KEYWORDS:

atrial fibrillation; mobile app; photoplethysmography; screening