Format

Send to

Choose Destination
Ann Surg. 2016 Oct;264(4):599-604. doi: 10.1097/SLA.0000000000001872.

The Role of Radio Frequency Detection System Embedded Surgical Sponges in Preventing Retained Surgical Sponges: A Prospective Evaluation in Patients Undergoing Emergency Surgery.

Author information

1
Los Angeles County Medical Center and University of Southern California, Division of Trauma Surgery and Surgical Critical Care, Los Angeles, CA.

Abstract

OBJECTIVE:

To prospectively evaluate the ability of radio frequency detection (RFD) system-embedded sponges to mitigate the incidence of retained surgical sponges (RSS) after emergency surgery.

BACKGROUND:

Emergency surgery patients are at high risk for retained foreign bodies.

METHODS:

All emergent trauma and nontrauma cavitary operations over a 5-year period (January 2010-December 2014) were prospectively enrolled. For damage-control procedures, only the definitive closure was included. RFD sponges were used exclusively throughout the study period. Before closure, the sponge and instrument count was followed by RFD scanning and x-ray evaluation for retained sponges. RSS and near-misses averted using the RFD system were analyzed.

RESULTS:

In all, 2051 patients [median (range)], aged 41 (1-101) years, 72.2% male, 46.8% trauma patients, underwent 2148 operations (1824 laparotomy, 100 thoracotomy, 30 sternotomy, and 97 combined). RFD detected retained sponges in 11 (0.5%) patients (81.8%laparotomy, 18.2% sternotomy) before cavitary closure. All postclosure x-rays were negative. No retained sponges were missed by the RFD system. Body mass index was 29 (23-43), estimated blood loss 1.0 L (0-23), and operating room time 160 minutes (71-869). Procedures started after 18:00 to 06:00 hours in 45.5% of the patients. The sponge count was incorrect in 36.4%, not performed due to time constraints in 45.5%, and correct in 18.2%. The additional cost of using RFD-embedded disposables was $0.17 for a 4X18 laparotomy sponge and $0.46 for a 10 pack of 12ply, 4X8.

CONCLUSIONS:

Emergent surgical procedures are high-risk for retained sponges, even when sponge counts are performed and found to be correct. Implementation of a RFD system was effective in preventing this complication and should be considered for emergent operations in an effort to improve patient safety.

PMID:
27433911
DOI:
10.1097/SLA.0000000000001872
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Wolters Kluwer
Loading ...
Support Center