Anal Biochem. 1989 Feb 1;176(2):360-4.

Spectrophotometric and high-performance chromatographic assays of hydroperoxides by the iodometric technique.

Author information

1
School of Biological Sciences, Macquarie University, Sydney, New South Wales, Australia.

Abstract

Samples containing between 1 and 50 nmol of hydroperoxides in oxygen-free methanol-acetic acid containing I- were incubated at 50 degrees C. This ensured reduction of all hydroperoxides tested in 15 min. Addition of sufficient Cd acetate to combine with the remaining I- allowed reading of absorbance of the I3- produced in open cuvettes. The absorbance was stable for several hours. It had a maximum at 358 nm with molar extinction coefficient of 2.97 X 10(4) 1 mol-1 cm-1. Sensitivity of the assay could be improved by injecting the I3- into a C-18 HPLC column eluted with methanol-water-acetic acid solution. Both methods are potentially suitable for assay of hydroperoxides in a wide range of biological materials.

PMID:
2742126
DOI:
10.1016/0003-2697(89)90323-0