Radiother Oncol. 2016 Apr;119(1):71-6. doi: 10.1016/j.radonc.2016.02.017. Epub 2016 Feb 27.

Uncertainties in estimating heart doses from 2D-tangential breast cancer radiotherapy.

Author information

1
Institute of Clinical Research, University of Southern Denmark, Odense, Denmark; Laboratory of Radiation Physics, Odense University Hospital, Denmark.
2
Nuffield Department of Population Health, University of Oxford, United Kingdom.
3
Institute of Clinical Research, University of Southern Denmark, Odense, Denmark; Department of Oncology, Odense University Hospital, Denmark.

Abstract

BACKGROUND AND PURPOSE:

We evaluated the accuracy of three methods of estimating radiation dose to the heart from two-dimensional tangential radiotherapy for breast cancer, as used in Denmark during 1982-2002.

MATERIAL AND METHODS:

Three tangential radiotherapy regimens were reconstructed using CT-based planning scans for 40 patients with left-sided and 10 with right-sided breast cancer. Setup errors and organ motion were simulated using estimated uncertainties. For left-sided patients, mean heart dose was related to maximum heart distance in the medial field.

RESULTS:

For left-sided breast cancer, mean heart dose estimated from individual CT-scans varied from <1Gy to >8Gy, and maximum dose from 5 to 50Gy for all three regimens, so that estimates based only on regimen had substantial uncertainty. When maximum heart distance was taken into account, the uncertainty was reduced and was comparable to the uncertainty of estimates based on individual CT-scans. For right-sided breast cancer patients, mean heart dose based on individual CT-scans was always <1Gy and maximum dose always <5Gy for all three regimens.

CONCLUSIONS:

The use of stored individual simulator films provides a method for estimating heart doses in left-tangential radiotherapy for breast cancer that is almost as accurate as estimates based on individual CT-scans.

KEYWORDS:

Breast cancer; Heart disease; Heart dose