Analysis of iodine in food samples by inductively coupled plasma-mass spectrometry

Food Addit Contam Part A Chem Anal Control Expo Risk Assess. 2016;33(2):282-90. doi: 10.1080/19440049.2015.1131337. Epub 2016 Jan 11.

Abstract

This work shows a method for the determination of iodine in a variety of food samples and reference materials using inductively coupled plasma-mass spectrometry (ICP-MS) following alkaline extraction. Optimisation of the addition of organic carbon showed that a minimum of 3% 2-propanol was necessary for a constant ratio of iodine to internal standard. The limit of quantification (LOQ), calculated as 30σ for the method, was 36 ng g(-1) in solid food samples. For method validation, seven standard reference materials (SRM) and 21 fortified food samples were used. The precision (%RSD) of the measurements was in the 2-7% range. Accuracies for the SRMs were 85-105%, while the fortified food samples showed 81-119% recoveries, including a number of samples fortified at 50% of the LOQ.

Keywords: ICP-MS; Iodine; alkaline extraction; carbon addition; reference materials; validation.

MeSH terms

  • Food Analysis / methods*
  • Iodine / analysis*
  • Mass Spectrometry

Substances

  • Iodine