J Orthop Surg (Hong Kong). 2015 Dec;23(3):336-40.

Drilling the near cortex with elongated figure-of-8 holes to reduce the stiffness of a locking compression plate construct.

Author information

1
Singapore General Hospital, Singapore.
2
Nanyang Technological University, Singapore.

Abstract

PURPOSE:

To compare the stiffness of locking compression plate (LCP) constructs with or without drilling the near cortex with elongated figure-of-8 holes.

METHODS:

24 synthetic bones were sawn to create a 10-mm gap and were fixed with a 9-hole 4.5-mm narrow LCP. In 12 bones, the near cortex of the adjacent holes to the LCP holes was drilled to create elongated figure-of-8 holes before screw insertion. The stiffness of LCP constructs under axial loading or 4-point bending was assessed by (1) dynamic quasi-physiological testing for fatigue strength, (2) quasi-static testing for stiffness, and (3) testing for absolute strength to failure.

RESULTS:

None of the 24 constructs had subcatastrophic or catastrophic failure after 10 000 cycles of fatigue loading (p=1.000). The axial stiffness reduced by 16% from 613±62 to 517±44 N/mm (p=0.012) in the case group, whereas the bending stiffness was 16±1 Nm2 in both groups (p=1.000). The maximum axial load to catastrophic failure was 1596±84 N for the control group and 1627±48 N for the case group (p=0.486), whereas the maximum bending moment to catastrophic failure was 79±12 and 80±10 Nm, respectively (p=0.919).

CONCLUSION:

Drilling the near cortex with elongated figure-of-8 holes reduces the axial stiffness of the LCP construct, without compromising its bending stiffness or strength.

KEYWORDS:

bone plates; compressive strength; fracture fixation

PMID:
26715713
DOI:
10.1177/230949901502300316
[Indexed for MEDLINE]