Label-free optical-resolution photoacoustic microscopy of superficial microvasculature using a compact visible laser diode excitation

Opt Express. 2015 Nov 30;23(24):31026-33. doi: 10.1364/OE.23.031026.

Abstract

We have developed laser-diode-based optical-resolution photoacoustic microscopy (LD-OR-PAM) of superficial microvasculature which has the desirable properties of being compact, low-cost, and label-free. A 300-mW visible pulsed laser diode was operated at a 405 ± 5 nm wavelength with a pulse energy as low as 52 nJ. By using a 3.6 MHz ultrasound transducer, the system was tested on carbon fibers with a lateral resolution of 0.95 µm and an SNR of 38 dB. The subcutaneous microvasculature on a mouse back was imaged without an exogenous contrast agent which demonstrates the potential of the proposed prototype for skin chromophores. Our eventual goal is to offer a practical and affordable multi-wavelength functional LD-OR-PAM instrument suitable for clinical applications.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Equipment Design
  • Equipment Failure Analysis
  • Feasibility Studies
  • Image Enhancement / instrumentation*
  • Lasers, Dye*
  • Lighting / instrumentation*
  • Mice
  • Microscopy / instrumentation*
  • Microvessels / cytology*
  • Miniaturization
  • Photoacoustic Techniques / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity