Determination of impurities in (124)I samples by high resolution gamma spectrometry

Appl Radiat Isot. 2016 Mar:109:261-263. doi: 10.1016/j.apradiso.2015.11.060. Epub 2015 Nov 25.

Abstract

(124)I is a radionuclide used in the diagnosis of tumors. The National Health Agency requires identification and activity measurement of impurities. Using gamma spectrometry with an efficiency calibrated high-purity germanium detector, impurities (125)I and (126)I in an (1)(24)I production sample were identified. Activity ratios of (125)I and (126)I to (124)I were approximately 0.5% and 98%, respectively.

Keywords: (124)I; Gamma-ray spectrometry; Impurities determination.