Distance-of-Flight Mass Spectrometry with IonCCD Detection and an Inductively Coupled Plasma Source

J Am Soc Mass Spectrom. 2016 Mar;27(3):371-9. doi: 10.1007/s13361-015-1295-7. Epub 2015 Nov 9.

Abstract

Distance-of-flight mass spectrometry (DOFMS) is demonstrated for the first time with a commercially available ion detector-the IonCCD camera. Because DOFMS is a velocity-based MS technique that provides spatially dispersive, simultaneous mass spectrometry, a position-sensitive ion detector is needed for mass-spectral collection. The IonCCD camera is a 5.1-cm long, 1-D array that is capable of simultaneous, multichannel ion detection along a focal plane, which makes it an attractive option for DOFMS. In the current study, the IonCCD camera is evaluated for DOFMS with an inductively coupled plasma (ICP) ionization source over a relatively short field-free mass-separation distance of 25.3-30.4 cm. The combination of ICP-DOFMS and the IonCCD detector results in a mass-spectral resolving power (FWHM) of approximately 900 and isotope-ratio precision equivalent to or slightly better than current ICP-TOFMS systems. The measured isotope-ratio precision in % relative standard deviation (%RSD) was ≥0.008%RSD for nonconsecutive isotopes at 10-ppm concentration (near the ion-signal saturation point) and ≥0.02%RSD for all isotopes at 1-ppm. Results of DOFMS with the IonCCD camera are also compared with those of two previously characterized detection setups.

Keywords: Distance-of-flight; Inductively coupled plasma; Instrumentation; IonCCD camera.