Format

Send to

Choose Destination
Cardiol Young. 2016 Mar;26(3):619-22. doi: 10.1017/S1047951115001730. Epub 2015 Aug 27.

Stent graft implantation in an aortic pseudoaneurysm associated with a fractured Cheatham-Platinum stent in aortic coarctation.

Author information

1
1Department of Radiology,University Medical Centre Ljubljana,Ljubljana,Slovenia.
2
2Department of Paediatrics,University Medical Centre Ljubljana,Ljubljana,Slovenia.

Abstract

We report a case of aortic pseudoaneurysm associated with a fractured bare Cheatham-Platinum stent following stenting for aortic coarctation. These complications were recognised 6 years after the implantation procedure and were successfully managed by percutaneous stent graft implantation. Staged approach for stent dilatation might prevent development of aortic pseudoaneurysms. In addition, careful follow-up is warranted after stenting for aortic coarctation, particularly in patients with recognised aortic wall injury.

KEYWORDS:

Aortic pseudoaneurysm; covered stent; stent fracture